A high-performance liquid chromatography method with fluorescence detection was developed for the quantification of riboflavin (RF), flavin mononucleotide (FMN), flavin adenine dinucleotide (FAD) and their photodegradation products, lumichrome (LC) and lumiflavin (LF), in liquid milk and milk products. Both sample preparation and chromatographic separation were studied to avoid acidic conditions that proved to affect flavin stability and degrade FAD into FMN. The sample preparation includes centrifugal skimming and ultrafiltration steps and is suitable for routine application. Linear response was obtained for individual flavins in the respective concentration ranges of interest and relative standard deviation (RSD) was lower than 5%, except for FAD (RSD 11%). The recovery ranged between 80–100%. The proposed method proved to be suitable for assessing flavins in commercial liquid milk and fermented milk products, and for monitoring the degradation of FAD, FMN and RF and the formation of LF and LC in bottled milk exposed to light during shelf storage.
Development of a HPLC method for the simultaneous analysis of riboflavin and other flavin compounds in liquid milk and milk products / D. Fracassetti, S. Limbo, P. D’Incecco, A. Tirelli, L. Pellegrino. - In: EUROPEAN FOOD RESEARCH AND TECHNOLOGY. - ISSN 1438-2377. - 244:9(2018 Sep), pp. 1545-1554.
Development of a HPLC method for the simultaneous analysis of riboflavin and other flavin compounds in liquid milk and milk products
D. Fracassetti
Primo
;S. LimboSecondo
;P. D’InceccoPenultimo
;A. Tirelli;L. PellegrinoUltimo
2018
Abstract
A high-performance liquid chromatography method with fluorescence detection was developed for the quantification of riboflavin (RF), flavin mononucleotide (FMN), flavin adenine dinucleotide (FAD) and their photodegradation products, lumichrome (LC) and lumiflavin (LF), in liquid milk and milk products. Both sample preparation and chromatographic separation were studied to avoid acidic conditions that proved to affect flavin stability and degrade FAD into FMN. The sample preparation includes centrifugal skimming and ultrafiltration steps and is suitable for routine application. Linear response was obtained for individual flavins in the respective concentration ranges of interest and relative standard deviation (RSD) was lower than 5%, except for FAD (RSD 11%). The recovery ranged between 80–100%. The proposed method proved to be suitable for assessing flavins in commercial liquid milk and fermented milk products, and for monitoring the degradation of FAD, FMN and RF and the formation of LF and LC in bottled milk exposed to light during shelf storage.File | Dimensione | Formato | |
---|---|---|---|
Fracassetti_et_al-2018-European_Food_Research_and_Technology.pdf
accesso riservato
Tipologia:
Publisher's version/PDF
Dimensione
897.05 kB
Formato
Adobe PDF
|
897.05 kB | Adobe PDF | Visualizza/Apri Richiedi una copia |
RF Manuscript_2018 03 14_rev3.pdf
Open Access dal 21/11/2019
Descrizione: Articolo principale
Tipologia:
Post-print, accepted manuscript ecc. (versione accettata dall'editore)
Dimensione
625.32 kB
Formato
Adobe PDF
|
625.32 kB | Adobe PDF | Visualizza/Apri |
Figures.pdf
accesso riservato
Descrizione: Figure dei dati sperimentali
Tipologia:
Altro
Dimensione
206.48 kB
Formato
Adobe PDF
|
206.48 kB | Adobe PDF | Visualizza/Apri Richiedi una copia |
Pubblicazioni consigliate
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.