Capillary zone electrophoresis (CZE) is a relatively new serum protein electrophoresis method with higher resolution than other electrophoretic techniques. Hypercholesterolemic dogs exhibit a peculiar CZE pattern. Specifically, they have a shoulder or peak immediately next to the albumin peak. We investigated the prevalence of this spurious peak in hypercholesterolemic dogs and its correlation with the serum cholesterol concentration. Moreover, possible discrepancies between the CZE and spectrophotometric (bromocresol green [BCG] method) albumin concentrations in those animals were evaluated, as well as the accuracy in measuring albumin by a different CZE fractionation system. We retrospectively enrolled 500 hypercholesterolemic and normotriglyceridemic dogs. Each electrophoretic curve was inspected visually to identify a spurious peak (prevalence of 68.8%). We chose 120 dogs to further investigate the albumin concentration; CZE albumin was significantly higher than measured using the BCG method. A weak but significant correlation (r = 0.412; p < 0.0001) was observed between the magnitude of the spurious peak and the serum cholesterol concentration. Finally, the significant difference between CZE and BCG albumin measurement disappeared (p = 0.92) when the spurious peak was considered as alpha 1-globulins instead of albumin.
Spurious capillary zone electrophoresis pattern in hypercholesterolemic dogs / G. Mangiagalli, S. Meazzi, A. Giordano, S. Rossi. - In: JOURNAL OF VETERINARY DIAGNOSTIC INVESTIGATION. - ISSN 1040-6387. - 35:2(2023), pp. 116-123. [10.1177/10406387221141872]
Spurious capillary zone electrophoresis pattern in hypercholesterolemic dogs
S. MeazziSecondo
;A. Giordano
Penultimo
;
2023
Abstract
Capillary zone electrophoresis (CZE) is a relatively new serum protein electrophoresis method with higher resolution than other electrophoretic techniques. Hypercholesterolemic dogs exhibit a peculiar CZE pattern. Specifically, they have a shoulder or peak immediately next to the albumin peak. We investigated the prevalence of this spurious peak in hypercholesterolemic dogs and its correlation with the serum cholesterol concentration. Moreover, possible discrepancies between the CZE and spectrophotometric (bromocresol green [BCG] method) albumin concentrations in those animals were evaluated, as well as the accuracy in measuring albumin by a different CZE fractionation system. We retrospectively enrolled 500 hypercholesterolemic and normotriglyceridemic dogs. Each electrophoretic curve was inspected visually to identify a spurious peak (prevalence of 68.8%). We chose 120 dogs to further investigate the albumin concentration; CZE albumin was significantly higher than measured using the BCG method. A weak but significant correlation (r = 0.412; p < 0.0001) was observed between the magnitude of the spurious peak and the serum cholesterol concentration. Finally, the significant difference between CZE and BCG albumin measurement disappeared (p = 0.92) when the spurious peak was considered as alpha 1-globulins instead of albumin.File | Dimensione | Formato | |
---|---|---|---|
Mangiagalli et al., JVDI 2023_bozza post referaggio.pdf
accesso aperto
Descrizione: Article
Tipologia:
Post-print, accepted manuscript ecc. (versione accettata dall'editore)
Dimensione
321.97 kB
Formato
Adobe PDF
|
321.97 kB | Adobe PDF | Visualizza/Apri |
10406387221141872.pdf
accesso riservato
Descrizione: Article
Tipologia:
Publisher's version/PDF
Dimensione
610.24 kB
Formato
Adobe PDF
|
610.24 kB | Adobe PDF | Visualizza/Apri Richiedi una copia |
Pubblicazioni consigliate
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.