Aims The emerging role of coronary computed tomography angiography (CCTA) as a non-invasive tool for atherosclerosis evaluation is supported by data reporting a good correlation between CCTA and intravascular ultrasound (IVUS) for plaque volume quantification. Aim of the present study was to evaluate whether a last generation CTscanner may improve coronary plaque volume assessment using IVUS as standard-of-reference. Methods and results From a registry of 1915 consecutive, all-comers, patients who underwent a clinically indicated IVUS evaluation we enrolled 59 patients who underwent CCTA with a 64-slice CT (Group 1) and 59 patients who underwent CCTA with whole-heart coverage CT scanner (Group 2). Patients who underwent CCTA with unfavourable heart rhythm were not excluded from the analysis. Image quality (4-point Likert scale) focused on plaque analysis was evaluated. Plaque volume quantification by CCTA was compared to IVUS. No difference in clinical characteristics was found between Group 1 and Group 2. Plaque volume quantification by CCTA was considered not feasible in 11 plaques of Group 1 and in 4 plaques of Group 2 (P = 0.09). Higher correlation for plaque volume quantification by CCTA vs. IVUS was demonstrated in Group 2 when compared with Group 1 (r = 0.9888 vs. 0.9499; P < 0.0001). The Bland–Altman analysis showed plaque volume overestimation by CCTA of 11.9 mm3 in Group 1 and 4 mm2 in Group 2 (P < 0.001). Effective radiation dose of CCTA was significantly lower in Group 2 vs. Group 1 (2.7 ± 0.9 vs. 8.1 ± 3.6 mSv, respectively; P < 0.001). Conclusions CCTA using a new scanner generation showed to be an accurate non-invasive tool to assess and quantify coronary plaque volume.

Plaque quantification by coronary computed tomography angiography using intravascular ultrasound as a reference standard: a comparison between standard and last generation computed tomography scanners / E. Conte, S. Mushtaq, G. Pontone, L. Li Piani, P. Ravagnani, S. Galli, C. Collet, J. Sonck, L. Di Odoardo, M. Guglielmo, A. Baggiano, D. Trabattoni, A. Annoni, M.E. Mancini, A. Formenti, G. Muscogiuri, M. Magatelli, F. Nicoli, C. Poggi, C. Fiorentini, A.L. Bartorelli, M. Pepi, P. Montorsi, D. Andreini. - In: EUROPEAN HEART JOURNAL. CARDIOVASCULAR IMAGING. - ISSN 2047-2404. - (2019 May 15). [Epub ahead of print] [10.1093/ehjci/jez089]

Plaque quantification by coronary computed tomography angiography using intravascular ultrasound as a reference standard: a comparison between standard and last generation computed tomography scanners

E. Conte
Primo
;
S. Mushtaq
Secondo
;
G. Pontone;S. Galli;L. Di Odoardo;A. Baggiano;A. Formenti;F. Nicoli;C. Fiorentini;A.L. Bartorelli;P. Montorsi;D. Andreini
2019

Abstract

Aims The emerging role of coronary computed tomography angiography (CCTA) as a non-invasive tool for atherosclerosis evaluation is supported by data reporting a good correlation between CCTA and intravascular ultrasound (IVUS) for plaque volume quantification. Aim of the present study was to evaluate whether a last generation CTscanner may improve coronary plaque volume assessment using IVUS as standard-of-reference. Methods and results From a registry of 1915 consecutive, all-comers, patients who underwent a clinically indicated IVUS evaluation we enrolled 59 patients who underwent CCTA with a 64-slice CT (Group 1) and 59 patients who underwent CCTA with whole-heart coverage CT scanner (Group 2). Patients who underwent CCTA with unfavourable heart rhythm were not excluded from the analysis. Image quality (4-point Likert scale) focused on plaque analysis was evaluated. Plaque volume quantification by CCTA was compared to IVUS. No difference in clinical characteristics was found between Group 1 and Group 2. Plaque volume quantification by CCTA was considered not feasible in 11 plaques of Group 1 and in 4 plaques of Group 2 (P = 0.09). Higher correlation for plaque volume quantification by CCTA vs. IVUS was demonstrated in Group 2 when compared with Group 1 (r = 0.9888 vs. 0.9499; P < 0.0001). The Bland–Altman analysis showed plaque volume overestimation by CCTA of 11.9 mm3 in Group 1 and 4 mm2 in Group 2 (P < 0.001). Effective radiation dose of CCTA was significantly lower in Group 2 vs. Group 1 (2.7 ± 0.9 vs. 8.1 ± 3.6 mSv, respectively; P < 0.001). Conclusions CCTA using a new scanner generation showed to be an accurate non-invasive tool to assess and quantify coronary plaque volume.
atherosclerosis; coronary computed tomography angiography; coronary plaque analysis; coronary plaque volume; intravascular ultrasound; whole-heart coverage CT
Settore MED/11 - Malattie dell'Apparato Cardiovascolare
Settore MED/36 - Diagnostica per Immagini e Radioterapia
15-mag-2019
Article (author)
File in questo prodotto:
File Dimensione Formato  
Plaque.pdf

accesso riservato

Tipologia: Publisher's version/PDF
Dimensione 1.03 MB
Formato Adobe PDF
1.03 MB Adobe PDF   Visualizza/Apri   Richiedi una copia