Introduction: Laboratory diagnosis of von Willebrand disease (VWD) is made by the measurement of von Willebrand factor (VWF) protein level and its activities. Current VWF activity tests include ristocetin cofactor and collagen binding (VWF:CB) assays. Aim: We have undertaken an evaluation of a new fully automated VWF:CB assay relative to an established enzyme-linked immunosorbent assay (ELISA) method. Methods: The two analytical systems operate with different detection principles: a chemiluminescent method performed on ACL AcuStar Analyzer (the former) and a colorimetric ELISA by Asserachrom Stago (the latter) (type III collagen from human placenta). The HemosIL AcuStar VWF:CB assay is a chemiluminescent 2-step immunoassay that uses magnetic particles coated with a type III collagen triple-helical peptide. VWF:CB levels were determined in 50 healthy subjects and 100 VWD patients (22 type 1, 73 type 2 and 5 type 3). Results: Eleven VWD samples reported VWF:CB values below the lower detection limit of one or both methods. The new method showed a good correlation with the ELISA method (r >.9, mean bias 3.85 IU/dL) in both healthy and VWD samples. One of 150 samples gave inconsistent results using the two assays, leading to an uncertain diagnosis of VWD type 1 (ELISA method) or type 2 MCB (fully automated method). Conclusion: The new assay is rapid and simple to use, with its ready-to-use reagent cartridges. This VWF:CB assay, in addition to the measurement of VWF:Ag and VWF:RCo made on the same platform, gives additional information for the diagnosis of VWD in both nonspecialized and reference laboratories.

A comparative evaluation of a new fully automated assay for von Willebrand factor collagen binding activity to an established method / F. Stufano, L. Baronciani, D. Mane-Padros, G. Cozzi, S. Faraudo, F. Peyvandi. - In: HAEMOPHILIA. - ISSN 1351-8216. - 24:1(2018 Jan), pp. 156-161.

A comparative evaluation of a new fully automated assay for von Willebrand factor collagen binding activity to an established method

L. Baronciani;F. Peyvandi
2018

Abstract

Introduction: Laboratory diagnosis of von Willebrand disease (VWD) is made by the measurement of von Willebrand factor (VWF) protein level and its activities. Current VWF activity tests include ristocetin cofactor and collagen binding (VWF:CB) assays. Aim: We have undertaken an evaluation of a new fully automated VWF:CB assay relative to an established enzyme-linked immunosorbent assay (ELISA) method. Methods: The two analytical systems operate with different detection principles: a chemiluminescent method performed on ACL AcuStar Analyzer (the former) and a colorimetric ELISA by Asserachrom Stago (the latter) (type III collagen from human placenta). The HemosIL AcuStar VWF:CB assay is a chemiluminescent 2-step immunoassay that uses magnetic particles coated with a type III collagen triple-helical peptide. VWF:CB levels were determined in 50 healthy subjects and 100 VWD patients (22 type 1, 73 type 2 and 5 type 3). Results: Eleven VWD samples reported VWF:CB values below the lower detection limit of one or both methods. The new method showed a good correlation with the ELISA method (r >.9, mean bias 3.85 IU/dL) in both healthy and VWD samples. One of 150 samples gave inconsistent results using the two assays, leading to an uncertain diagnosis of VWD type 1 (ELISA method) or type 2 MCB (fully automated method). Conclusion: The new assay is rapid and simple to use, with its ready-to-use reagent cartridges. This VWF:CB assay, in addition to the measurement of VWF:Ag and VWF:RCo made on the same platform, gives additional information for the diagnosis of VWD in both nonspecialized and reference laboratories.
English
von Willebrand disease; von Willebrand factor; VWD; VWF; VWF:CB; hematology; genetics (clinical)
Settore MED/09 - Medicina Interna
Articolo
Esperti anonimi
Ricerca di base
Pubblicazione scientifica
gen-2018
Wiley Blackwell Publishing
24
1
156
161
6
Pubblicato
Periodico con rilevanza internazionale
scopus
pubmed
crossref
Aderisco
info:eu-repo/semantics/article
A comparative evaluation of a new fully automated assay for von Willebrand factor collagen binding activity to an established method / F. Stufano, L. Baronciani, D. Mane-Padros, G. Cozzi, S. Faraudo, F. Peyvandi. - In: HAEMOPHILIA. - ISSN 1351-8216. - 24:1(2018 Jan), pp. 156-161.
reserved
Prodotti della ricerca::01 - Articolo su periodico
6
262
Article (author)
si
F. Stufano, L. Baronciani, D. Mane-Padros, G. Cozzi, S. Faraudo, F. Peyvandi
File in questo prodotto:
File Dimensione Formato  
Stufano_et_al-2018-Haemophilia.pdf

accesso riservato

Tipologia: Publisher's version/PDF
Dimensione 531.13 kB
Formato Adobe PDF
531.13 kB Adobe PDF   Visualizza/Apri   Richiedi una copia
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/557177
Citazioni
  • ???jsp.display-item.citation.pmc??? 2
  • Scopus 6
  • ???jsp.display-item.citation.isi??? 6
social impact