Background The use of glycated albumin (GA) has been proposed as an additional glycemic control marker particularly useful in intermediate-term monitoring and in situation when HbA1c test is not reliable. Methods We have performed the first multicenter evaluation of the analytical performance of the enzymatic method quantILab Glycated Albumin assay implemented on the most widely used clinical chemistry analyzers (i.e. Abbott Architect C8000, Beckman Coulter AU 480 and 680, Roche Cobas C6000, Siemens ADVIA 2400 and 2400 XPT). Results The repeatability of the GA measurement (expressed as CV, %) implemented in the participating centers ranged between 0.9% and 1.2%. The within-laboratory CVs ranged between 1.2% and 1.6%. A good alignment between laboratories was found, with correlation coefficients from 0.996 to 0.998. Linearity was confirmed in the range from 7.6 to 84.7%. Conclusion The new enzymatic method for glycated albumin evaluated by our investigation is suitable for clinical use.

Multicenter evaluation of an enzymatic method for glycated albumin / R. Paleari, G. Bonetti, C. Callà, M. Carta, F. Ceriotti, N. Di Gaetano, M. Ferri, E. Guerra, G. Lavalle, C.L. Cascio, F.G. Martino, M. Montagnana, M. Moretti, G. Santini, D. Scribano, R. Testa, A. Vero, A. Mosca. - In: CLINICA CHIMICA ACTA. - ISSN 0009-8981. - 469(2017), pp. 81-86.

Multicenter evaluation of an enzymatic method for glycated albumin

R. Paleari
Primo
;
A. Mosca
2017

Abstract

Background The use of glycated albumin (GA) has been proposed as an additional glycemic control marker particularly useful in intermediate-term monitoring and in situation when HbA1c test is not reliable. Methods We have performed the first multicenter evaluation of the analytical performance of the enzymatic method quantILab Glycated Albumin assay implemented on the most widely used clinical chemistry analyzers (i.e. Abbott Architect C8000, Beckman Coulter AU 480 and 680, Roche Cobas C6000, Siemens ADVIA 2400 and 2400 XPT). Results The repeatability of the GA measurement (expressed as CV, %) implemented in the participating centers ranged between 0.9% and 1.2%. The within-laboratory CVs ranged between 1.2% and 1.6%. A good alignment between laboratories was found, with correlation coefficients from 0.996 to 0.998. Linearity was confirmed in the range from 7.6 to 84.7%. Conclusion The new enzymatic method for glycated albumin evaluated by our investigation is suitable for clinical use.
Diabetes; Glycated albumin; Multicenter evaluation; Biochemistry; Clinical Biochemistry; Biochemistry (medical)
Settore BIO/12 - Biochimica Clinica e Biologia Molecolare Clinica
2017
Article (author)
File in questo prodotto:
File Dimensione Formato  
2017_Glycated albumin CCA.pdf

accesso riservato

Tipologia: Publisher's version/PDF
Dimensione 626.61 kB
Formato Adobe PDF
626.61 kB Adobe PDF   Visualizza/Apri   Richiedi una copia
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/505219
Citazioni
  • ???jsp.display-item.citation.pmc??? 5
  • Scopus 20
  • ???jsp.display-item.citation.isi??? 13
social impact