BACKGROUND: The performances of 2 noninvasive CPAP systems (high flow and low flow air-entrainment masks) were compared to the Boussignac valve in 3 different scenarios. METHODS: Scenario 1: pneumatic lung simulator with a tachypnea pattern (tidal volume 800 mL at 40 breaths/min). Scenario 2: Ten healthy subjects studied during tidal breaths and tachypnea. Scenario 3: Twenty ICU subjects enrolled for a noninvasive CPAP session. Differences between set and effective CPAP level and F(IO(2)), as well as the lowest airway pressure and the pressure swing around the imposed CPAP level, were analyzed. The lowest airway pressure and swing were correlated to the pressure-time product (area of the airway pressure curve below the CPAP level) measured with the simulator. P(aO(2)) was a subject's further performance index. RESULTS: Lung simulator: Boussignac F(IO(2)) was 0.54, even if supplied with pure oxygen. The air-entrainment masks had higher swing than the Boussignac (P = .007). Pressure-time product correlated better with pressure swing (Spearman correlation coefficient [ρ] = 0.97) than with lowest airway pressure (ρ = 0.92). In healthy subjects, the high-flow air-entrainment mask showed lower difference between set and effective F(IO(2)) (P < .001), and lowest airway pressure (P < .001), compared to the Boussignac valve. In all measurements the Boussignac valve showed higher than imposed CPAP level (P < .001). In ICU subjects the high-flow mask had lower swing than the Boussignac valve (P = .03) with similar P(aO(2)) increase. CONCLUSIONS: High-flow air-entrainment mask showed the best performance in human subjects. During high flow demand, the Boussignac valve delivered lower than expected F(IO(2)) and showed higher dynamic hyper-pressurization than the air-entrainment masks.

Noninvasive CPAP with face mask : comparison among new air-entrainment masks and the Boussignac Valve / G. Mistraletti, M. Giacomini, G. Sabbatini, R. Pinciroli, E. Mantovani, M. Umbrello, D. Palmisano, P. Formenti, A. Destrebecq, G. Iapichino. - In: RESPIRATORY CARE. - ISSN 0020-1324. - 58:2(2013 Feb), pp. 305-312. [10.4187/respcare.01598]

Noninvasive CPAP with face mask : comparison among new air-entrainment masks and the Boussignac Valve

G. Mistraletti
Primo
;
G. Sabbatini;E. Mantovani;M. Umbrello;D. Palmisano;P. Formenti;A. Destrebecq
Penultimo
;
G. Iapichino
Ultimo
2013

Abstract

BACKGROUND: The performances of 2 noninvasive CPAP systems (high flow and low flow air-entrainment masks) were compared to the Boussignac valve in 3 different scenarios. METHODS: Scenario 1: pneumatic lung simulator with a tachypnea pattern (tidal volume 800 mL at 40 breaths/min). Scenario 2: Ten healthy subjects studied during tidal breaths and tachypnea. Scenario 3: Twenty ICU subjects enrolled for a noninvasive CPAP session. Differences between set and effective CPAP level and F(IO(2)), as well as the lowest airway pressure and the pressure swing around the imposed CPAP level, were analyzed. The lowest airway pressure and swing were correlated to the pressure-time product (area of the airway pressure curve below the CPAP level) measured with the simulator. P(aO(2)) was a subject's further performance index. RESULTS: Lung simulator: Boussignac F(IO(2)) was 0.54, even if supplied with pure oxygen. The air-entrainment masks had higher swing than the Boussignac (P = .007). Pressure-time product correlated better with pressure swing (Spearman correlation coefficient [ρ] = 0.97) than with lowest airway pressure (ρ = 0.92). In healthy subjects, the high-flow air-entrainment mask showed lower difference between set and effective F(IO(2)) (P < .001), and lowest airway pressure (P < .001), compared to the Boussignac valve. In all measurements the Boussignac valve showed higher than imposed CPAP level (P < .001). In ICU subjects the high-flow mask had lower swing than the Boussignac valve (P = .03) with similar P(aO(2)) increase. CONCLUSIONS: High-flow air-entrainment mask showed the best performance in human subjects. During high flow demand, the Boussignac valve delivered lower than expected F(IO(2)) and showed higher dynamic hyper-pressurization than the air-entrainment masks.
CPAP; Development and evaluation/technology assessment; Devices; Equipment; Noninvasive ventilation; PEEP
Settore MED/41 - Anestesiologia
feb-2013
Article (author)
File in questo prodotto:
File Dimensione Formato  
RC 2013 Mistraletti - CPAP Ventumask vs Boussignac.pdf

accesso riservato

Tipologia: Publisher's version/PDF
Dimensione 195.92 kB
Formato Adobe PDF
195.92 kB Adobe PDF   Visualizza/Apri   Richiedi una copia
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/193941
Citazioni
  • ???jsp.display-item.citation.pmc??? 2
  • Scopus 9
  • ???jsp.display-item.citation.isi??? 7
social impact