Background: Patients suspected of platelet function defects represent a diagnostic challenge for the clinical laboratory, mainly due to the complexity and poor standardization of screening methods. We compared a new flow-based chip-equipped point-of-care (T-TAS) device with lumi-aggregometry and other specific tests. Materials and methods: The study included 96 patients suspected of platelet function defects and 26 patients referred to hospital for an evaluation of residual platelet function while on antiplatelet therapy. Results: Forty-eight of 96 patients displayed abnormal platelet function by lumi-aggregometry, and 10 of them had defective granule content and were classified as δ-storage pool disease (δ-SPD). T-TAS compared favorably with lumi-aggregometry in detecting the most severe forms of platelet function defects (i.e., δ-SPD) [test agreement (lumi-light transmission aggregometry [lumi-LTA] vs T-TAS) for the δ-SPD subgroup was 80% and K CHOEN 0.695. T-TAS was less sensitive to milder platelet function defects (i.e., primary secretion defects [PSD]). Concerning patients on antiplatelets, test agreement (lumi-LTA vs T-TAS) in detecting patients who were responders to this therapy was 54%; K CHOEN 0.150. Discussion: The results indicate that T-TAS can detect the more severe forms of platelet function defects such as δ-SPD. There is limited agreement of T-TAS with lumi-aggregometry in identifying responders to antiplatelets. However, this poor agreement is commonly shared by lumi-aggregometry and other devices owing to the lack of test specificity and of prospective data from clinical trials linking platelet function with therapeutic efficacy.

Flow-chamber device (T-TAS) to diagnose patients suspected of platelet function defects / A. Lecchi, S. La Marca, L. Padovan, M. Boscarino, F. Peyvandi, A. Tripodi. - In: BLOOD TRANSFUSION. - ISSN 2385-2070. - 22:1(2024 Feb), pp. 55-64. [10.2450/2023.0208-22]

Flow-chamber device (T-TAS) to diagnose patients suspected of platelet function defects

M. Boscarino;F. Peyvandi
Penultimo
;
A. Tripodi
Ultimo
2024

Abstract

Background: Patients suspected of platelet function defects represent a diagnostic challenge for the clinical laboratory, mainly due to the complexity and poor standardization of screening methods. We compared a new flow-based chip-equipped point-of-care (T-TAS) device with lumi-aggregometry and other specific tests. Materials and methods: The study included 96 patients suspected of platelet function defects and 26 patients referred to hospital for an evaluation of residual platelet function while on antiplatelet therapy. Results: Forty-eight of 96 patients displayed abnormal platelet function by lumi-aggregometry, and 10 of them had defective granule content and were classified as δ-storage pool disease (δ-SPD). T-TAS compared favorably with lumi-aggregometry in detecting the most severe forms of platelet function defects (i.e., δ-SPD) [test agreement (lumi-light transmission aggregometry [lumi-LTA] vs T-TAS) for the δ-SPD subgroup was 80% and K CHOEN 0.695. T-TAS was less sensitive to milder platelet function defects (i.e., primary secretion defects [PSD]). Concerning patients on antiplatelets, test agreement (lumi-LTA vs T-TAS) in detecting patients who were responders to this therapy was 54%; K CHOEN 0.150. Discussion: The results indicate that T-TAS can detect the more severe forms of platelet function defects such as δ-SPD. There is limited agreement of T-TAS with lumi-aggregometry in identifying responders to antiplatelets. However, this poor agreement is commonly shared by lumi-aggregometry and other devices owing to the lack of test specificity and of prospective data from clinical trials linking platelet function with therapeutic efficacy.
Settore MED/09 - Medicina Interna
feb-2024
Article (author)
File in questo prodotto:
File Dimensione Formato  
Lecchi BLOOD TRANSF 2.pdf

accesso riservato

Tipologia: Publisher's version/PDF
Dimensione 784.9 kB
Formato Adobe PDF
784.9 kB Adobe PDF   Visualizza/Apri   Richiedi una copia
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/1078088
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 8
  • ???jsp.display-item.citation.isi??? 7
  • OpenAlex ND
social impact