The present pilot study was designed by hypothesizing a possible correlation between lack of accuracy in implant placement and peri-implant hard and soft tissue health. A total of five patients underwent computer-guided implant surgery and full-arch immediate loading between 2013 and 2014. They subsequently underwent postoperative cone-beam computed tomography (CBCT). After a follow-up of 5 years, all patients were recalled for a clinical-radiographic evaluation of peri-implant health status. The mean linear deviation was 0.5 ± 0.2 mm at the implant’s head and 0.6 ± 0.2 mm at the implant’s apex, while the mean angular deviation of the long axis was 2.8° ± 1.2°. A mean marginal bone loss (MBL) of 1.16 ± 0.94 mm and 2.01 ± 1.76 mm was observed after 1 and 5 years of follow-up, respectively. At 5 years, the mean peri-implant probing depth (PPD) was 4.09 ± 1.44 mm, 66.6% of the evaluated implants showed peri-implant bleeding on probing (BOP), keratinized mucosa (KM) was <2 mm in 48.4% of cases, and mucosal recession (REC) ≥ 1 mm was assessed in 45.4% of the included implants. A negative correlation was observed between bucco-palatal/lingual linear inaccuracy and MBL, PPD, BOP, and KM.

Correlation between Accuracy in Computer-Guided Implantology and Peri-Implant Tissue Stability: A Prospective Clinical and Radiological Pilot Study / P.P. Poli, M. Manfredini, C. Maiorana, F.E. Salina, M. Beretta. - In: JOURNAL OF CLINICAL MEDICINE. - ISSN 2077-0383. - 12:15(2023 Aug 03), pp. 5098.1-5098.15. [10.3390/jcm12155098]

Correlation between Accuracy in Computer-Guided Implantology and Peri-Implant Tissue Stability: A Prospective Clinical and Radiological Pilot Study

P.P. Poli
Primo
;
M. Manfredini
Secondo
;
C. Maiorana;M. Beretta
Ultimo
2023

Abstract

The present pilot study was designed by hypothesizing a possible correlation between lack of accuracy in implant placement and peri-implant hard and soft tissue health. A total of five patients underwent computer-guided implant surgery and full-arch immediate loading between 2013 and 2014. They subsequently underwent postoperative cone-beam computed tomography (CBCT). After a follow-up of 5 years, all patients were recalled for a clinical-radiographic evaluation of peri-implant health status. The mean linear deviation was 0.5 ± 0.2 mm at the implant’s head and 0.6 ± 0.2 mm at the implant’s apex, while the mean angular deviation of the long axis was 2.8° ± 1.2°. A mean marginal bone loss (MBL) of 1.16 ± 0.94 mm and 2.01 ± 1.76 mm was observed after 1 and 5 years of follow-up, respectively. At 5 years, the mean peri-implant probing depth (PPD) was 4.09 ± 1.44 mm, 66.6% of the evaluated implants showed peri-implant bleeding on probing (BOP), keratinized mucosa (KM) was <2 mm in 48.4% of cases, and mucosal recession (REC) ≥ 1 mm was assessed in 45.4% of the included implants. A negative correlation was observed between bucco-palatal/lingual linear inaccuracy and MBL, PPD, BOP, and KM.
computer-guided implantology; dental implants; digital dentistry; peri-implant hard tissue; peri-implant health; peri-implant soft tissue;
Settore MED/28 - Malattie Odontostomatologiche
3-ago-2023
Article (author)
File in questo prodotto:
File Dimensione Formato  
jcm-12-05098.pdf

accesso aperto

Tipologia: Publisher's version/PDF
Dimensione 4.08 MB
Formato Adobe PDF
4.08 MB Adobe PDF Visualizza/Apri