Objective: Little is known about the cardiovascular changes after TEVAR and regarding the impact on aortic stiffness for different stent graft generations specifically, following changes in device design. The present study evaluated the stent graft induced aortic stiffening of two generations of the Valiant thoracic aortic stent graft.Methods: This was an ex vivo porcine investigation using an experimental mock circulatory loop. Thoracic aortas of young healthy pigs were harvested and connected to the mock circulatory loop. At a 60 bpm heart rate and stable mean arterial pressure, baseline aortic characteristics were obtained. Pulse wave velocity (PWV) was calculated before and after stent graft deployment. Paired and independent sample t tests or their nonparametric alternatives were performed to test for differences where appropriate.Results: Twenty porcine thoracic aortas were divided into two equal subgroups, in which a Valiant Captivia or a Valiant Navion stent graft was deployed. Both stent grafts were similar in diameter and length. Baseline aortic characteristics did not differ between the subgroups. Mean arterial pressure values did not change after either stent graft, while pulse pressures increased statistically significantly after Captivia (mean 44 +/- 10 mmHg to 51 +/- 13 mmHg, p = .002) but not after Navion. Mean baseline PWV increased after both Captivia (4.4 +/- 0.6 m/s to 4.8 +/- 0.7 m/s, p = .007) and Navion (4.6 +/- 0.7 m/s to 4.9 +/- 0.7 m/s, p = .002). There was no statistically significant difference in the mean percentage increase in PWV for either subgroup (8 +/- 4% vs. 6 +/- 4%, p = .25).Conclusion: These experimental findings showed no statistically significant difference in the percentage increase of aortic PWV after either stent graft generation and confirm that TEVAR increases aortic PWV. As a surrogate for aortic stiffness, this calls for further improvements in future thoracic aortic stent graft designs regarding device compliance. (c) 2023 The Author(s). Published by Elsevier Ltd on behalf of European Society for Vascular Surgery. This is an open access article under the CC BY license (http://creativecommons.org/licenses/by/4.0/).

Comparison of Two Generations of Thoracic Aortic Stent Grafts and Their Impact on Aortic Stiffness in an Ex Vivo Porcine Model / T.J. Mandigers, M. Conti, S. Allievi, F. Dedola, D. Bissacco, D. Bianchi, S. Marconi, M. Domanin, J.A. Van Herwaarden, F. Auricchio, S. Trimarchi. - In: EJVES VASCULAR FORUM. - ISSN 2666-688X. - 59:(2023), pp. 8-14. [10.1016/j.ejvsvf.2023.04.001]

Comparison of Two Generations of Thoracic Aortic Stent Grafts and Their Impact on Aortic Stiffness in an Ex Vivo Porcine Model

S. Allievi;D. Bissacco;M. Domanin;S. Trimarchi
Ultimo
2023

Abstract

Objective: Little is known about the cardiovascular changes after TEVAR and regarding the impact on aortic stiffness for different stent graft generations specifically, following changes in device design. The present study evaluated the stent graft induced aortic stiffening of two generations of the Valiant thoracic aortic stent graft.Methods: This was an ex vivo porcine investigation using an experimental mock circulatory loop. Thoracic aortas of young healthy pigs were harvested and connected to the mock circulatory loop. At a 60 bpm heart rate and stable mean arterial pressure, baseline aortic characteristics were obtained. Pulse wave velocity (PWV) was calculated before and after stent graft deployment. Paired and independent sample t tests or their nonparametric alternatives were performed to test for differences where appropriate.Results: Twenty porcine thoracic aortas were divided into two equal subgroups, in which a Valiant Captivia or a Valiant Navion stent graft was deployed. Both stent grafts were similar in diameter and length. Baseline aortic characteristics did not differ between the subgroups. Mean arterial pressure values did not change after either stent graft, while pulse pressures increased statistically significantly after Captivia (mean 44 +/- 10 mmHg to 51 +/- 13 mmHg, p = .002) but not after Navion. Mean baseline PWV increased after both Captivia (4.4 +/- 0.6 m/s to 4.8 +/- 0.7 m/s, p = .007) and Navion (4.6 +/- 0.7 m/s to 4.9 +/- 0.7 m/s, p = .002). There was no statistically significant difference in the mean percentage increase in PWV for either subgroup (8 +/- 4% vs. 6 +/- 4%, p = .25).Conclusion: These experimental findings showed no statistically significant difference in the percentage increase of aortic PWV after either stent graft generation and confirm that TEVAR increases aortic PWV. As a surrogate for aortic stiffness, this calls for further improvements in future thoracic aortic stent graft designs regarding device compliance. (c) 2023 The Author(s). Published by Elsevier Ltd on behalf of European Society for Vascular Surgery. This is an open access article under the CC BY license (http://creativecommons.org/licenses/by/4.0/).
English
Aortic stiffness; Captivia; Experimental investigation; Navion; Pulse wave velocity; Valiant;
Settore MED/22 - Chirurgia Vascolare
Articolo
Esperti anonimi
Pubblicazione scientifica
2023
Elsevier
59
8
14
7
Pubblicato
Periodico con rilevanza internazionale
pubmed
wos
scopus
crossref
Aderisco
info:eu-repo/semantics/article
Comparison of Two Generations of Thoracic Aortic Stent Grafts and Their Impact on Aortic Stiffness in an Ex Vivo Porcine Model / T.J. Mandigers, M. Conti, S. Allievi, F. Dedola, D. Bissacco, D. Bianchi, S. Marconi, M. Domanin, J.A. Van Herwaarden, F. Auricchio, S. Trimarchi. - In: EJVES VASCULAR FORUM. - ISSN 2666-688X. - 59:(2023), pp. 8-14. [10.1016/j.ejvsvf.2023.04.001]
open
Prodotti della ricerca::01 - Articolo su periodico
11
262
Article (author)
Periodico senza Impact Factor
T.J. Mandigers, M. Conti, S. Allievi, F. Dedola, D. Bissacco, D. Bianchi, S. Marconi, M. Domanin, J.A. Van Herwaarden, F. Auricchio, S. Trimarchi
File in questo prodotto:
File Dimensione Formato  
main (1).pdf

accesso aperto

Tipologia: Publisher's version/PDF
Dimensione 713.12 kB
Formato Adobe PDF
713.12 kB Adobe PDF Visualizza/Apri
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/994608
Citazioni
  • ???jsp.display-item.citation.pmc??? 0
  • Scopus 3
  • ???jsp.display-item.citation.isi??? 1
social impact