Numerical simulations of thoracic endovascular aortic repair (TEVAR) may be implemented in the preoperative workflow if credible and reliable. We present the application of a TEVAR simulation methodology to an 82-year-old woman with a penetrating atherosclerotic ulcer in the left hemiarch, that underwent a left common carotid artery to left subclavian artery bypass and consequent TEVAR in zone 2. During the intervention, kinking of the distal thoracic stent graft occurred and the simulation was able to reproduce this event. This report highlights the potential and reliability of TEVAR simulations to predict perioperative adverse events and short-term postoperative technical results.

Utilizing numerical simulations to prevent stent graft kinking during thoracic endovascular aortic repair / T.J. Mandigers, A. Ramella, D. Bissacco, M. Domanin, J.A. van Herwaarden, G. Luraghi, F. Migliavacca, S. Trimarchi. - In: JOURNAL OF VASCULAR SURGERY CASES AND INNOVATIVE TECHNIQUES. - ISSN 2468-4287. - 9:3(2023 Sep), pp. 101269.1-101269.4. [10.1016/j.jvscit.2023.101269]

Utilizing numerical simulations to prevent stent graft kinking during thoracic endovascular aortic repair

D. Bissacco;M. Domanin;S. Trimarchi
Ultimo
2023

Abstract

Numerical simulations of thoracic endovascular aortic repair (TEVAR) may be implemented in the preoperative workflow if credible and reliable. We present the application of a TEVAR simulation methodology to an 82-year-old woman with a penetrating atherosclerotic ulcer in the left hemiarch, that underwent a left common carotid artery to left subclavian artery bypass and consequent TEVAR in zone 2. During the intervention, kinking of the distal thoracic stent graft occurred and the simulation was able to reproduce this event. This report highlights the potential and reliability of TEVAR simulations to predict perioperative adverse events and short-term postoperative technical results.
Computational modeling; Finite element analysis; Patient-specific in silico model; TEVAR; Thoracic endovascular aortic repair;
Settore MED/22 - Chirurgia Vascolare
set-2023
Article (author)
File in questo prodotto:
File Dimensione Formato  
1-s2.0-S2468428723001788-main(1).pdf

accesso aperto

Tipologia: Publisher's version/PDF
Dimensione 1.25 MB
Formato Adobe PDF
1.25 MB Adobe PDF Visualizza/Apri
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/994582
Citazioni
  • ???jsp.display-item.citation.pmc??? 0
  • Scopus 2
  • ???jsp.display-item.citation.isi??? 1
social impact