In this contribution we report on further investigations of the recently-evaluated Constant-Fraction Time-over-Threshold (CF-ToT) method for neutron/gamma-ray Pulse Shape Discrimination (PSD). The superiority of the CF-ToT PSD method over the constant-threshold (CT-ToT) method was previously demonstrated, down to low neutron energy thresholds of 100 keVee. Here, we report on a quantitative comparison between the traditionally used Charge Comparison (CC) method and the CF-ToT method using a stilbene scintillator coupled to a silicon photomultiplier, implementing an offline analysis of recorded fast-neutron and gamma-ray waveforms. An optimization of the constant fraction value indicates that a 20%-fraction yields the optimum figure-of-merit (FOM) and gamma-ray peak-to-valley (P/V) ratio. The results obtained for a particle energy threshold of 100 keVee (kilo electron Volt electron equivalent) show that the FOM and P/V values achieved with the CF-ToT method are superior to those obtained using the standard CC method. In addition, a first electronic implementation of the CF-ToT method was performed using simple circuitry suitable for multichannel architecture. Initial results obtained with this circuit prototype are presented.
Optimization and first electronic implementation of the Constant-Fraction Time-Over-Threshold pulse shape discrimination method / A. Roy, D. Vartsky, I. Mor, C. Boiano, S. Brambilla, S. Riboldi, E.O. Cohen, Y. Yehuda-Zada, A.B.A.L. Arazi. - In: JOURNAL OF INSTRUMENTATION. - ISSN 1748-0221. - (2023), pp. 18 P05022.1-18 P05022.12. [Epub ahead of print] [10.1088/1748-0221/18/05/P05022]
Optimization and first electronic implementation of the Constant-Fraction Time-Over-Threshold pulse shape discrimination method
S. Riboldi;
2023
Abstract
In this contribution we report on further investigations of the recently-evaluated Constant-Fraction Time-over-Threshold (CF-ToT) method for neutron/gamma-ray Pulse Shape Discrimination (PSD). The superiority of the CF-ToT PSD method over the constant-threshold (CT-ToT) method was previously demonstrated, down to low neutron energy thresholds of 100 keVee. Here, we report on a quantitative comparison between the traditionally used Charge Comparison (CC) method and the CF-ToT method using a stilbene scintillator coupled to a silicon photomultiplier, implementing an offline analysis of recorded fast-neutron and gamma-ray waveforms. An optimization of the constant fraction value indicates that a 20%-fraction yields the optimum figure-of-merit (FOM) and gamma-ray peak-to-valley (P/V) ratio. The results obtained for a particle energy threshold of 100 keVee (kilo electron Volt electron equivalent) show that the FOM and P/V values achieved with the CF-ToT method are superior to those obtained using the standard CC method. In addition, a first electronic implementation of the CF-ToT method was performed using simple circuitry suitable for multichannel architecture. Initial results obtained with this circuit prototype are presented.File | Dimensione | Formato | |
---|---|---|---|
Roy_2023_J._Inst._18_P05022.pdf
accesso aperto
Tipologia:
Publisher's version/PDF
Dimensione
2.64 MB
Formato
Adobe PDF
|
2.64 MB | Adobe PDF | Visualizza/Apri |
Pubblicazioni consigliate
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.