Background: Ventricular fibrillation (VF) waveform analysis has been proposed as a potential non-invasive guide to optimize timing of defibrillation. Methods: The AMplitude Spectrum Area (AMSA) trial is an open-label, multicenter randomized controlled study reporting the first in-human use of AMSA analysis in out-of-hospital cardiac arrest (OHCA). The primary efficacy endpoint was the termination of VF for an AMSA ≥ 15.5 mV-Hz. Adult shockable OHCAs randomly received either an AMSA-guided cardiopulmonary resuscitation (CPR) or a standard-CPR. Randomization and allocation to trial group were carried out centrally. In the AMSA-guided CPR, an initial AMSA ≥ 15.5 mV-Hz prompted for immediate defibrillation, while lower values favored chest compression (CC). After completion of the first 2-min CPR cycle, an AMSA < 6.5 mV-Hz deferred defibrillation in favor of an additional 2-min CPR cycle. AMSA was measured and displayed in real-time during CC pauses for ventilation with a modified defibrillator. Findings: The trial was early discontinued for low recruitment due to the COVID-19 pandemics. A total of 31 patients were recruited in 3 Italian cities, 19 in AMSA-CPR and 12 in standard-CPR, and included in the data analysis. No difference in primary outcome was observed between the two groups. Termination of VF occurred in 74% of patients in the AMSA-CPR compared to 75% in the standard CPR (OR 0.93 [95% CI 0.18-4.90]). No adverse events were reported. Interpretation: AMSA was used prospectively in human patients during ongoing CPR. In this small trial, an AMSA-guided defibrillation provided no evidence of an improvement in termination of VF. Trial registration: NCT03237910. Funding: European Commission - Horizon 2020; ZOLL Medical Corp., Chelmsford, USA (unrestricted grant); Italian Ministry of Health - Current research IRCCS.

Amplitude Spectrum Area of ventricular fibrillation to guide defibrillation: a small open-label, pseudo-randomized controlled multicenter trial / L. Ruggeri, F. Fumagalli, F. Bernasconi, F. Semeraro, J.M.T.A. Meessen, A. Blanda, M. Migliari, A. Magliocca, G. Gordini, R. Fumagalli, G. Sechi, A. Pesenti, M.B. Skrifvars, Y. Li, R. Latini, L. Wik, G. Ristagno. - In: EBIOMEDICINE. - ISSN 2352-3964. - 90:(2023 Apr), pp. 104544.1-104544.11. [10.1016/j.ebiom.2023.104544]

Amplitude Spectrum Area of ventricular fibrillation to guide defibrillation: a small open-label, pseudo-randomized controlled multicenter trial

A. Magliocca;A. Pesenti;G. Ristagno
Ultimo
2023

Abstract

Background: Ventricular fibrillation (VF) waveform analysis has been proposed as a potential non-invasive guide to optimize timing of defibrillation. Methods: The AMplitude Spectrum Area (AMSA) trial is an open-label, multicenter randomized controlled study reporting the first in-human use of AMSA analysis in out-of-hospital cardiac arrest (OHCA). The primary efficacy endpoint was the termination of VF for an AMSA ≥ 15.5 mV-Hz. Adult shockable OHCAs randomly received either an AMSA-guided cardiopulmonary resuscitation (CPR) or a standard-CPR. Randomization and allocation to trial group were carried out centrally. In the AMSA-guided CPR, an initial AMSA ≥ 15.5 mV-Hz prompted for immediate defibrillation, while lower values favored chest compression (CC). After completion of the first 2-min CPR cycle, an AMSA < 6.5 mV-Hz deferred defibrillation in favor of an additional 2-min CPR cycle. AMSA was measured and displayed in real-time during CC pauses for ventilation with a modified defibrillator. Findings: The trial was early discontinued for low recruitment due to the COVID-19 pandemics. A total of 31 patients were recruited in 3 Italian cities, 19 in AMSA-CPR and 12 in standard-CPR, and included in the data analysis. No difference in primary outcome was observed between the two groups. Termination of VF occurred in 74% of patients in the AMSA-CPR compared to 75% in the standard CPR (OR 0.93 [95% CI 0.18-4.90]). No adverse events were reported. Interpretation: AMSA was used prospectively in human patients during ongoing CPR. In this small trial, an AMSA-guided defibrillation provided no evidence of an improvement in termination of VF. Trial registration: NCT03237910. Funding: European Commission - Horizon 2020; ZOLL Medical Corp., Chelmsford, USA (unrestricted grant); Italian Ministry of Health - Current research IRCCS.
Amplitude spectrum area; Cardiac arrest; Defibrillation; Ventricular fibrillation; Waveform analysis
Settore MED/41 - Anestesiologia
   European Sudden Cardiac Arrest network: towards Prevention, Education and NEw Treatment
   ESCAPE-NET
   European Commission
   Horizon 2020 Framework Programme
   733381
apr-2023
Article (author)
File in questo prodotto:
File Dimensione Formato  
1-s2.0-S2352396423001093-main.pdf

accesso aperto

Descrizione: Article
Tipologia: Publisher's version/PDF
Dimensione 492.45 kB
Formato Adobe PDF
492.45 kB Adobe PDF Visualizza/Apri
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/970358
Citazioni
  • ???jsp.display-item.citation.pmc??? 2
  • Scopus 6
  • ???jsp.display-item.citation.isi??? 5
social impact