(1) Background: Recently, novel dental implants that are characterized by different levels of surface roughness in the distinct parts of the fixture’s body have been introduced in the market. These surface characteristics could affect the primary stability of the implants. The aim of this in vitro study was to compare the primary stability of short and long implants, characterized by multiscale surface roughness, inserted on polyurethane blocks. The secondary aim was to understand if the implant length could be a crucial factor in the decision-making in immediate or rather than delayed loading protocol in the different bone densities. (2) Methods: A total of 20 cylindrical dental implants with a diameter of 5.0 mm were tested for the lengths 6.0 mm (short implants) versus 13.0 mm (long implants) on two different solid rigid polyurethane blocks (20 and 30 PCF). The primary stability was evaluated by measuring the insertion torque value (ITV), the removal torque (RTV), and the resonance frequency analysis RFA. (3) Results: The values of ITV, RTV, and RFA showed the same trend in all measurements. Long implants showed a significantly higher primary stability on 30 PCF blocks that present mechanical properties similar to high-density bone. On the contrary, no relevant differences were found on 20 PCF blocks, which mimic trabecular bone density. (4) Conclusions: The impact of fixture length on the primary stability of implants with multiscale surface roughness is significant in 30 PCF polyurethane corresponding to higher bone density, but not in lower ones.

Fixture Length and Primary Stability: An In Vitro Study on Polyurethane Foam / M. Petrini, M. Tumedei, A. Cipollina, S. D'Ercole, M.S. Di Carmine, A. Piattelli, M. Re, G. Iezzi. - In: APPLIED SCIENCES. - ISSN 2076-3417. - 12:5(2022), pp. 2683.1-2683.9. [10.3390/app12052683]

Fixture Length and Primary Stability: An In Vitro Study on Polyurethane Foam

M. Tumedei
Secondo
;
2022

Abstract

(1) Background: Recently, novel dental implants that are characterized by different levels of surface roughness in the distinct parts of the fixture’s body have been introduced in the market. These surface characteristics could affect the primary stability of the implants. The aim of this in vitro study was to compare the primary stability of short and long implants, characterized by multiscale surface roughness, inserted on polyurethane blocks. The secondary aim was to understand if the implant length could be a crucial factor in the decision-making in immediate or rather than delayed loading protocol in the different bone densities. (2) Methods: A total of 20 cylindrical dental implants with a diameter of 5.0 mm were tested for the lengths 6.0 mm (short implants) versus 13.0 mm (long implants) on two different solid rigid polyurethane blocks (20 and 30 PCF). The primary stability was evaluated by measuring the insertion torque value (ITV), the removal torque (RTV), and the resonance frequency analysis RFA. (3) Results: The values of ITV, RTV, and RFA showed the same trend in all measurements. Long implants showed a significantly higher primary stability on 30 PCF blocks that present mechanical properties similar to high-density bone. On the contrary, no relevant differences were found on 20 PCF blocks, which mimic trabecular bone density. (4) Conclusions: The impact of fixture length on the primary stability of implants with multiscale surface roughness is significant in 30 PCF polyurethane corresponding to higher bone density, but not in lower ones.
Insertion torque; Polyurethane foam; Primary stability; Resonance frequency
Settore MED/50 - Scienze Tecniche Mediche Applicate
Settore MED/28 - Malattie Odontostomatologiche
2022
Article (author)
File in questo prodotto:
File Dimensione Formato  
applsci-12-02683.pdf

accesso aperto

Descrizione: Article
Tipologia: Publisher's version/PDF
Dimensione 1.23 MB
Formato Adobe PDF
1.23 MB Adobe PDF Visualizza/Apri