Introduction: Unrecognized incomplete pulmonary vein (PV) isolation during the index procedure, can be a major cause of clinical recurrences of atrial fibrillation (AF) after cryoballoon (CB) ablation. We aimed to characterize the extension of the lesions produced by CB ablation and to assess the value of using an ultra-high resolution electroanatomic mapping (UHDM) system to detect incomplete CB lesions. Materials and methods: Twenty-nine consecutive patients from the CHARISMA registry undergoing AF ablation at four Italian centers were prospectively evaluated. The Rhythmia (TM) mapping system and the Orion (TM) (Boston Scientific) mapping catheter were used to systematically map the left atrium and PVs before and after cryoablation. Results: A total of 116 PVs were targeted and isolated. Quantitative assessment of the lesions revealed a significant reduction of the antral surface area of the PV, resulting in an ablated area of 5.7 & PLUSMN; 0.7 cm(2) and 5.1 & PLUSMN; 0.8 cm(2) for the left PV pair and right PV pair, respectively (p = 0.0068). The mean posterior wall (PW) area was 22.9 & PLUSMN; 2 cm(2) and, following PV isolation, 44.8 & PLUSMN; 6% of the PW area was ablated. After CB ablation, complete isolation of each PV was documented by the POLARMap (TM) catheter in all patients. By contrast, confirmatory UHDM and the Lumipoint (TM) tool unveiled PV signals in 1 out of 114 of the PVs (0.9%). Over 30-day follow-up, no major procedure-related adverse events were reported. After a mean follow-up of 333 days, 89.7% of patients were free from arrhythmia recurrence. Conclusion: The lesion extension achieved by the new CB ablation system involved the PV antrum, with less than 50% of the PW remaining untouched. The new system, with short tip and circular mapping catheter, failed to achieve PV isolation in only 0.9% of all PVs treated.

Ultra-high-resolution assessment of lesion extension after cryoballoon ablation for pulmonary vein isolation / F. Spera, M.L. Narducci, G. Bencardino, F. Perna, A. Bisignani, G. Pinnacchio, C. Tondo, R. Maggio, G. Stabile, S. Iacopino, F. Tundo, A. Ferraro, A. De Simone, M. Malacrida, F. Pintus, F. Crea, G. Pelargonio. - In: FRONTIERS IN CARDIOVASCULAR MEDICINE. - ISSN 2297-055X. - 9:(2022), pp. 985182.1-985182.10. [10.3389/fcvm.2022.985182]

Ultra-high-resolution assessment of lesion extension after cryoballoon ablation for pulmonary vein isolation

C. Tondo;
2022

Abstract

Introduction: Unrecognized incomplete pulmonary vein (PV) isolation during the index procedure, can be a major cause of clinical recurrences of atrial fibrillation (AF) after cryoballoon (CB) ablation. We aimed to characterize the extension of the lesions produced by CB ablation and to assess the value of using an ultra-high resolution electroanatomic mapping (UHDM) system to detect incomplete CB lesions. Materials and methods: Twenty-nine consecutive patients from the CHARISMA registry undergoing AF ablation at four Italian centers were prospectively evaluated. The Rhythmia (TM) mapping system and the Orion (TM) (Boston Scientific) mapping catheter were used to systematically map the left atrium and PVs before and after cryoablation. Results: A total of 116 PVs were targeted and isolated. Quantitative assessment of the lesions revealed a significant reduction of the antral surface area of the PV, resulting in an ablated area of 5.7 & PLUSMN; 0.7 cm(2) and 5.1 & PLUSMN; 0.8 cm(2) for the left PV pair and right PV pair, respectively (p = 0.0068). The mean posterior wall (PW) area was 22.9 & PLUSMN; 2 cm(2) and, following PV isolation, 44.8 & PLUSMN; 6% of the PW area was ablated. After CB ablation, complete isolation of each PV was documented by the POLARMap (TM) catheter in all patients. By contrast, confirmatory UHDM and the Lumipoint (TM) tool unveiled PV signals in 1 out of 114 of the PVs (0.9%). Over 30-day follow-up, no major procedure-related adverse events were reported. After a mean follow-up of 333 days, 89.7% of patients were free from arrhythmia recurrence. Conclusion: The lesion extension achieved by the new CB ablation system involved the PV antrum, with less than 50% of the PW remaining untouched. The new system, with short tip and circular mapping catheter, failed to achieve PV isolation in only 0.9% of all PVs treated.
ablation < electrophysiology; atrial fibrillation; cryoballoon ablation; electrograms; scar ablation
Settore MED/11 - Malattie dell'Apparato Cardiovascolare
2022
Article (author)
File in questo prodotto:
File Dimensione Formato  
Ultra high.pdf

accesso aperto

Tipologia: Publisher's version/PDF
Dimensione 1.75 MB
Formato Adobe PDF
1.75 MB Adobe PDF Visualizza/Apri
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/951352
Citazioni
  • ???jsp.display-item.citation.pmc??? 0
  • Scopus 7
  • ???jsp.display-item.citation.isi??? 5
  • OpenAlex ND
social impact