Left atrial appendage (LAA) closure may be complicated by occluder embolization. Percutaneous retrieval is preferred for devices embolized to large vessels. In this report, the successful percutaneous retrieval of an LAA occluder embolized to the abdominal aorta was followed by several complications, culminating in iatrogenic aortic rupture requiring endovascular repair. (Level of Difficulty: Advanced.).
Aortic Rupture and Hemorrhagic Shock After Percutaneous Retrieval of an Embolized Left Atrial Appendage Occluder / A. Mastrangelo, S. Galli, P. Montorsi, A.L. Bartorelli. - In: JACC. CASE REPORTS. - ISSN 2666-0849. - 4:8(2022 Apr 20), pp. 486-490. [10.1016/j.jaccas.2022.03.004]
Aortic Rupture and Hemorrhagic Shock After Percutaneous Retrieval of an Embolized Left Atrial Appendage Occluder
A. Mastrangelo
;P. Montorsi;A.L. Bartorelli
2022
Abstract
Left atrial appendage (LAA) closure may be complicated by occluder embolization. Percutaneous retrieval is preferred for devices embolized to large vessels. In this report, the successful percutaneous retrieval of an LAA occluder embolized to the abdominal aorta was followed by several complications, culminating in iatrogenic aortic rupture requiring endovascular repair. (Level of Difficulty: Advanced.).File | Dimensione | Formato | |
---|---|---|---|
35493795.pdf
accesso aperto
Tipologia:
Publisher's version/PDF
Dimensione
869.67 kB
Formato
Adobe PDF
|
869.67 kB | Adobe PDF | Visualizza/Apri |
Pubblicazioni consigliate