Axial flow pumps have gained increased acceptance in recent years as a bridge to heart transplantation and, more recently, as destination therapy. As left ventricular (LV) assist device dysfunction will be increasingly prevalent, the aim of our work was to introduce an echocardiographic management protocol as a guide to recognize the causes of pump failure. In this article we describe the echocardiographic approach to 5 episodes of malfunction of an axial flow pump (DeBakey, MicroMed Technology Inc, Houston, Tex) in 4 patients: 4 episodes caused by thrombosis of LV assist device and one caused by abnormal increase of systemic vascular resistance. In our experience, echocardiography played a pivotal role in clinical management of LV assist device failure. It allowed us to: assess patency and position of inflow and outflow cannulae; research the source of thromboembolic material; assess adequate LV filling and unloading; and optimize right ventricular function, volume replacement therapy, and pharmacologic support.

Left ventricular support by axial flow pump: the echocardiographic approach to device malfunction / E. Catena, F. Milazzo, E. Montorsi, G. Bruschi, A. Cannata, C. Russo, A. Barosi, G. Tarelli, P. Tartara, R. Paino, E. Vitali. - In: JOURNAL OF THE AMERICAN SOCIETY OF ECHOCARDIOGRAPHY. - ISSN 0894-7317. - 18:12(2005 Dec), pp. 1422.e7.1-1422.e7.7. [10.1016/j.echo.2005.04.011]

Left ventricular support by axial flow pump: the echocardiographic approach to device malfunction

E. Catena;E. Vitali
2005

Abstract

Axial flow pumps have gained increased acceptance in recent years as a bridge to heart transplantation and, more recently, as destination therapy. As left ventricular (LV) assist device dysfunction will be increasingly prevalent, the aim of our work was to introduce an echocardiographic management protocol as a guide to recognize the causes of pump failure. In this article we describe the echocardiographic approach to 5 episodes of malfunction of an axial flow pump (DeBakey, MicroMed Technology Inc, Houston, Tex) in 4 patients: 4 episodes caused by thrombosis of LV assist device and one caused by abnormal increase of systemic vascular resistance. In our experience, echocardiography played a pivotal role in clinical management of LV assist device failure. It allowed us to: assess patency and position of inflow and outflow cannulae; research the source of thromboembolic material; assess adequate LV filling and unloading; and optimize right ventricular function, volume replacement therapy, and pharmacologic support.
Echocardiography; Equipment Failure; Equipment Failure Analysis; Heart-Assist Devices; Humans; Ventricular Dysfunction, Left
Settore MED/41 - Anestesiologia
dic-2005
Article (author)
File in questo prodotto:
File Dimensione Formato  
1-s2.0-S0894731705003378-main (1).pdf

accesso aperto

Tipologia: Publisher's version/PDF
Dimensione 421.73 kB
Formato Adobe PDF
421.73 kB Adobe PDF Visualizza/Apri
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/939066
Citazioni
  • ???jsp.display-item.citation.pmc??? 3
  • Scopus 49
  • ???jsp.display-item.citation.isi??? 28
social impact