Objectives: The TOP-HOLE (Towards OPtimal glycoHemOgLobin tEsting) project aimed to validate the HbA1c enzymatic method on the Abbott Alinity c platform and to implement the HbA1c testing process on the total laboratory automation (TLA) system of our institution. Methods: Three different measuring systems were employed: Architect c4000 stand-alone (s-a), Alinity c s-a, and Alinity c TLA. Eight frozen whole blood samples, IFCC valueassigned, were used for checking trueness. A comparison study testing transferability of HbA1c results from Architect to Alinity was also performed. The alignment of Alinity TLA vs. s-a was verified and the measurement uncertainty (MU) estimated according to ISO 20914:2019. Turnaround time (TAT) and full time equivalent (FTE) were used as efficiency indicators. Results: For HbA1c concentrations covering cut-offs adopted in clinical setting, the bias for both Architect and Alinity s-a was negligible. When compared with Architect, Alinity showed a mean positive bias of 0.54 mmol/mol, corresponding to a mean difference of 0.87%. A perfect alignment of Alinity TLA to the Alinity s-a was shown, and a MU of 1.58% was obtained, widely fulfilling the desirable 3.0% goal. After the full automation of HbA1c testing, 90% of results were released with a maximum TAT of 1 h, 0.30 FTE resource was also saved. Conclusions: The traceability of Alinity HbA1c enzymatic assay to the IFCC reference system was correctly implemented. We successfully completed the integration of the HbA1c testing on our TLA system, without worsening the optimal analytical performance. The shift of HbA1c testing from s-a mode to TLA significantly decreased TAT

A step towards optimal efficiency of HbA1c measurement as a first-line laboratory test: the TOP-HOLE (Towards OPtimal glycoHemOgLobin tEsting) project / S. Pasqualetti, A. Carnevale, A. Dolci, M. Panteghini. - In: CLINICAL CHEMISTRY AND LABORATORY MEDICINE. - ISSN 1434-6621. - 60:3(2022 Feb 23), pp. 441-450. [10.1515/cclm-2021-1249]

A step towards optimal efficiency of HbA1c measurement as a first-line laboratory test: the TOP-HOLE (Towards OPtimal glycoHemOgLobin tEsting) project

A. Dolci
Penultimo
;
M. Panteghini
Ultimo
2022

Abstract

Objectives: The TOP-HOLE (Towards OPtimal glycoHemOgLobin tEsting) project aimed to validate the HbA1c enzymatic method on the Abbott Alinity c platform and to implement the HbA1c testing process on the total laboratory automation (TLA) system of our institution. Methods: Three different measuring systems were employed: Architect c4000 stand-alone (s-a), Alinity c s-a, and Alinity c TLA. Eight frozen whole blood samples, IFCC valueassigned, were used for checking trueness. A comparison study testing transferability of HbA1c results from Architect to Alinity was also performed. The alignment of Alinity TLA vs. s-a was verified and the measurement uncertainty (MU) estimated according to ISO 20914:2019. Turnaround time (TAT) and full time equivalent (FTE) were used as efficiency indicators. Results: For HbA1c concentrations covering cut-offs adopted in clinical setting, the bias for both Architect and Alinity s-a was negligible. When compared with Architect, Alinity showed a mean positive bias of 0.54 mmol/mol, corresponding to a mean difference of 0.87%. A perfect alignment of Alinity TLA to the Alinity s-a was shown, and a MU of 1.58% was obtained, widely fulfilling the desirable 3.0% goal. After the full automation of HbA1c testing, 90% of results were released with a maximum TAT of 1 h, 0.30 FTE resource was also saved. Conclusions: The traceability of Alinity HbA1c enzymatic assay to the IFCC reference system was correctly implemented. We successfully completed the integration of the HbA1c testing on our TLA system, without worsening the optimal analytical performance. The shift of HbA1c testing from s-a mode to TLA significantly decreased TAT
No
English
clinical governance; diabetes; HbA1c; metrological traceability; total laboratory automation
Settore BIO/12 - Biochimica Clinica e Biologia Molecolare Clinica
Articolo
Sì, ma tipo non specificato
Pubblicazione scientifica
23-feb-2022
gen-2022
De Gruyter
60
3
441
450
10
Pubblicato
Periodico con rilevanza internazionale
Riferibilità Metrologica in Medicina di Laboratorio (CIRME)
manual
Aderisco
info:eu-repo/semantics/article
A step towards optimal efficiency of HbA1c measurement as a first-line laboratory test: the TOP-HOLE (Towards OPtimal glycoHemOgLobin tEsting) project / S. Pasqualetti, A. Carnevale, A. Dolci, M. Panteghini. - In: CLINICAL CHEMISTRY AND LABORATORY MEDICINE. - ISSN 1434-6621. - 60:3(2022 Feb 23), pp. 441-450. [10.1515/cclm-2021-1249]
reserved
Prodotti della ricerca::01 - Articolo su periodico
4
262
Article (author)
no
S. Pasqualetti, A. Carnevale, A. Dolci, M. Panteghini
File in questo prodotto:
File Dimensione Formato  
cclm-2022_HbA1c TOP-HOLE project.pdf

accesso riservato

Tipologia: Publisher's version/PDF
Dimensione 1.05 MB
Formato Adobe PDF
1.05 MB Adobe PDF   Visualizza/Apri   Richiedi una copia
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/904867
Citazioni
  • ???jsp.display-item.citation.pmc??? 0
  • Scopus 1
  • ???jsp.display-item.citation.isi??? 1
social impact