The implant primary stability is a fundamental condition for avoiding implant micro-motions that might result in fibrous encapsulation; its achievement is facilitated by macro- and micro-geometry of the implant, and by the bone density and architecture at the intended implant site. The aim was to evaluate an osteocondensing drilling protocol for dental implant positioning compared to standard protocol on polyurethane block. A total of 40 implants, 20 for each osteocondensing group (Test) and 20 for standard drilling group (Control), were positioned. Insertion torque (IT), removal torque (RT), and Periotest were measured. A IT, RT, and Periotest significant difference was present in favor of the implant osteocondensing protocol. The results suggested that osteocondensing protocol represents a useful technique for implant placement in poor density bone.

Implant primary stability with an osteocondensation drilling protocol in different density polyurethane blocks / S. Fanali, M. Tumedei, P. Pignatelli, F. Inchingolo, P. Pennacchietti, G. Pace, A. Piattelli. - In: COMPUTER METHODS IN BIOMECHANICS AND BIOMEDICAL ENGINEERING. - ISSN 1025-5842. - 24:1(2021), pp. 14-20. [10.1080/10255842.2020.1806251]

Implant primary stability with an osteocondensation drilling protocol in different density polyurethane blocks

M. Tumedei
;
2021

Abstract

The implant primary stability is a fundamental condition for avoiding implant micro-motions that might result in fibrous encapsulation; its achievement is facilitated by macro- and micro-geometry of the implant, and by the bone density and architecture at the intended implant site. The aim was to evaluate an osteocondensing drilling protocol for dental implant positioning compared to standard protocol on polyurethane block. A total of 40 implants, 20 for each osteocondensing group (Test) and 20 for standard drilling group (Control), were positioned. Insertion torque (IT), removal torque (RT), and Periotest were measured. A IT, RT, and Periotest significant difference was present in favor of the implant osteocondensing protocol. The results suggested that osteocondensing protocol represents a useful technique for implant placement in poor density bone.
Bone condensation; implant osteotomy; polyurethane block; primary stability; Bone and Bones; Humans; Movement; Polyurethanes; Torque; Dental Implants
Settore MED/50 - Scienze Tecniche Mediche Applicate
Settore MED/28 - Malattie Odontostomatologiche
2021
Article (author)
File in questo prodotto:
File Dimensione Formato  
CM IN BIOM AND BIOMED ENGIN-Fanali 2020.pdf

accesso riservato

Tipologia: Publisher's version/PDF
Dimensione 1.76 MB
Formato Adobe PDF
1.76 MB Adobe PDF   Visualizza/Apri   Richiedi una copia
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/896687
Citazioni
  • ???jsp.display-item.citation.pmc??? 12
  • Scopus 75
  • ???jsp.display-item.citation.isi??? 66
social impact