The aim of this human cadaver study was to assess the accuracy of zygomatic/pterygoid implant placement using custom-made bone-supported laser sintered titanium templates. For this purpose, pre-surgical planning was done on computed tomography scans of each cadaver. Surgical guides were printed using direct metal laser sintering technology. Four zygomatic and two pterygoid implants were inserted in each case using the guided protocol and related tools. Post-operative computed tomography (CT) scans were obtained to evaluate deviations between the planned and inserted implants. Accuracy was measured by overlaying the real position in the post-operative CT on the virtual presurgical placement of the implant in a CT image. Descriptive and bivariate analyses of the data were performed. As a result, a total of 40 zygomatic and 20 pterygoid implants were inserted in 10 cadavers. The mean deviations between the planned and the placed zygomatic and pterygoid implants were respectively (mean ± SD): 1.69◦ ± 1.12◦ and 4.15◦ ± 3.53◦ for angular deviation. Linear distance deviations: 0.93 mm ± 1.23 mm and 1.35 mm ± 1.45 mm at platform depth, 1.35 mm ± 0.78 mm and 1.81 mm ± 1.47 mm at apical plane, 1.07 mm ± 1.47 mm and 1.22 mm ± 1.44 mm for apical depth. In conclusion, the surgical guide system showed accuracy for all the variables studied and allowed acceptable and accurate implant placement regardless of the case complexity.

A novel guided zygomatic and pterygoid implant surgery system: A human cadaver study on accuracy / F. Grecchi, L.V. Stefanelli, F. Grivetto, E. Grecchi, R. Siev, Z. Mazor, M. Del Fabbro, N. Pranno, A. Franchina, V. Di Lucia, F. De Angelis, F. Goker. - In: INTERNATIONAL JOURNAL OF ENVIRONMENTAL RESEARCH AND PUBLIC HEALTH. - ISSN 1661-7827. - 18:11(2021), pp. 6142.1-6142.20. [10.3390/ijerph18116142]

A novel guided zygomatic and pterygoid implant surgery system: A human cadaver study on accuracy

E. Grecchi;M. Del Fabbro;F. Goker
2021

Abstract

The aim of this human cadaver study was to assess the accuracy of zygomatic/pterygoid implant placement using custom-made bone-supported laser sintered titanium templates. For this purpose, pre-surgical planning was done on computed tomography scans of each cadaver. Surgical guides were printed using direct metal laser sintering technology. Four zygomatic and two pterygoid implants were inserted in each case using the guided protocol and related tools. Post-operative computed tomography (CT) scans were obtained to evaluate deviations between the planned and inserted implants. Accuracy was measured by overlaying the real position in the post-operative CT on the virtual presurgical placement of the implant in a CT image. Descriptive and bivariate analyses of the data were performed. As a result, a total of 40 zygomatic and 20 pterygoid implants were inserted in 10 cadavers. The mean deviations between the planned and the placed zygomatic and pterygoid implants were respectively (mean ± SD): 1.69◦ ± 1.12◦ and 4.15◦ ± 3.53◦ for angular deviation. Linear distance deviations: 0.93 mm ± 1.23 mm and 1.35 mm ± 1.45 mm at platform depth, 1.35 mm ± 0.78 mm and 1.81 mm ± 1.47 mm at apical plane, 1.07 mm ± 1.47 mm and 1.22 mm ± 1.44 mm for apical depth. In conclusion, the surgical guide system showed accuracy for all the variables studied and allowed acceptable and accurate implant placement regardless of the case complexity.
English
Accuracy; CAD/CAM; Computer aided implantology; Dynamic navigation; Guidance; Guided surgery; Navigation; Pterygoid implants; Surgical guides; Surgical templates; Zygomatic implant
Settore MED/50 - Scienze Tecniche Mediche Applicate
Settore MED/29 - Chirurgia Maxillofacciale
Settore MED/28 - Malattie Odontostomatologiche
Articolo
Esperti anonimi
Ricerca applicata
Pubblicazione scientifica
2021
MDPI AG
18
11
6142
1
20
20
Pubblicato
Periodico con rilevanza internazionale
scopus
pubmed
crossref
wos
Aderisco
info:eu-repo/semantics/article
A novel guided zygomatic and pterygoid implant surgery system: A human cadaver study on accuracy / F. Grecchi, L.V. Stefanelli, F. Grivetto, E. Grecchi, R. Siev, Z. Mazor, M. Del Fabbro, N. Pranno, A. Franchina, V. Di Lucia, F. De Angelis, F. Goker. - In: INTERNATIONAL JOURNAL OF ENVIRONMENTAL RESEARCH AND PUBLIC HEALTH. - ISSN 1661-7827. - 18:11(2021), pp. 6142.1-6142.20. [10.3390/ijerph18116142]
open
Prodotti della ricerca::01 - Articolo su periodico
12
262
Article (author)
Periodico senza Impact Factor
F. Grecchi, L.V. Stefanelli, F. Grivetto, E. Grecchi, R. Siev, Z. Mazor, M. Del Fabbro, N. Pranno, A. Franchina, V. Di Lucia, F. De Angelis, F. Goker...espandi
File in questo prodotto:
File Dimensione Formato  
ijerph-18-06142.pdf

accesso aperto

Tipologia: Publisher's version/PDF
Dimensione 9.9 MB
Formato Adobe PDF
9.9 MB Adobe PDF Visualizza/Apri
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/861125
Citazioni
  • ???jsp.display-item.citation.pmc??? 6
  • Scopus 18
  • ???jsp.display-item.citation.isi??? 18
  • OpenAlex 20
social impact