Background Recently, the Food and Drug Administration issued a recall for the subcutaneous implantable cardioverter-defibrillator (S-ICD) because of the possibility of lead ruptures and accelerated battery depletion. Objective The aim of this study was to evaluate device-related complications over time in a large real-world multicenter S-ICD cohort. Methods Patients implanted with an S-ICD from January 2015 to June 2020 were enrolled from a 19-institution European registry (Experience from the Long-term Italian S-ICD registry [ELISIR]; ClinicalTrials.gov identifier NCT0473876). Device-related complication rates over follow-up were collected. Last follow-up of patients was performed after the Boston Scientific recall issue. Results A total of 1254 patients (median age 52.0 [interquartile range 41.0–62.2] years; 77.6% men; 30.9% ischemic) was enrolled. Over a follow-up of 23.2 (12.8–37.8) months, complications were observed in 117 patients (9.3%) for a total of 127 device-related complications (23.6% managed conservatively and 76.4% required reintervention). Twenty-seven patients (2.2%) had unanticipated generator replacement after 3.6 (3.3–3.9) years, while 4 (0.3%) had lead rupture. Body mass index (hazard ratio [HR] 1.063 [95% confidence interval 1.028–1.100]; P = .000), chronic kidney disease (HR 1.960 [1.191–3.225]; P = .008), and oral anticoagulation (HR 1.437 [1.010–2.045]; P = .043) were associated with an increase in overall complications, whereas older age (HR 0.980 [0.967–0.994]; P = .007) and procedure performed in high-volume centers (HR 0.463 [0.300–0.715]; P = .001) were protective factors. Conclusion The overall complication rate over 23.2 months of follow-up in a multicenter S-ICD cohort was 9.3%. Early unanticipated device battery depletions occurred in 2.2% of patients, while lead fracture was observed in 0.3%, which is in line with the expected rates reported by Boston Scientific.

Long term complications in patients implanted with subcutaneous implantable defibrillators Real-world data from the Extended ELISIR experience / A. Gasperetti, M. Schiavone, M. Ziacchi, J. Vogler, A. Breitenstein, M. Laredo, P. Palmisano, D. Ricciardi, G. Mitacchione, P. Compagnucci, A. Bisignani, A. Angeletti, M. Casella, F. Picarelli, T. Fink, L. Kaiser, S. Hakmi, L. Calò, C. Pignalberi, L. Santini, C. Lavalle, E. Pisanò, I. Olivotto, C. Tondo, A. Curnis, A.D. Russo, N. Badenco, J. Steffel, C.J. Love, R. Tilz, G. Forleo, M. Biffi. - In: HEART RHYTHM. - ISSN 1547-5271. - (2021), pp. 1-9. [Epub ahead of print] [10.1016/j.hrthm.2021.07.008]

Long term complications in patients implanted with subcutaneous implantable defibrillators Real-world data from the Extended ELISIR experience

A. Gasperetti
Primo
;
M. Schiavone
Secondo
;
C. Tondo;
2021

Abstract

Background Recently, the Food and Drug Administration issued a recall for the subcutaneous implantable cardioverter-defibrillator (S-ICD) because of the possibility of lead ruptures and accelerated battery depletion. Objective The aim of this study was to evaluate device-related complications over time in a large real-world multicenter S-ICD cohort. Methods Patients implanted with an S-ICD from January 2015 to June 2020 were enrolled from a 19-institution European registry (Experience from the Long-term Italian S-ICD registry [ELISIR]; ClinicalTrials.gov identifier NCT0473876). Device-related complication rates over follow-up were collected. Last follow-up of patients was performed after the Boston Scientific recall issue. Results A total of 1254 patients (median age 52.0 [interquartile range 41.0–62.2] years; 77.6% men; 30.9% ischemic) was enrolled. Over a follow-up of 23.2 (12.8–37.8) months, complications were observed in 117 patients (9.3%) for a total of 127 device-related complications (23.6% managed conservatively and 76.4% required reintervention). Twenty-seven patients (2.2%) had unanticipated generator replacement after 3.6 (3.3–3.9) years, while 4 (0.3%) had lead rupture. Body mass index (hazard ratio [HR] 1.063 [95% confidence interval 1.028–1.100]; P = .000), chronic kidney disease (HR 1.960 [1.191–3.225]; P = .008), and oral anticoagulation (HR 1.437 [1.010–2.045]; P = .043) were associated with an increase in overall complications, whereas older age (HR 0.980 [0.967–0.994]; P = .007) and procedure performed in high-volume centers (HR 0.463 [0.300–0.715]; P = .001) were protective factors. Conclusion The overall complication rate over 23.2 months of follow-up in a multicenter S-ICD cohort was 9.3%. Early unanticipated device battery depletions occurred in 2.2% of patients, while lead fracture was observed in 0.3%, which is in line with the expected rates reported by Boston Scientific.
English
S-ICD; device complications; device recall; lead complications; sub-cutaneous implantable cardioverter defibrillator;
Settore MED/11 - Malattie dell'Apparato Cardiovascolare
Articolo
Esperti anonimi
Pubblicazione scientifica
2021
14-lug-2021
Elsevier
1
9
9
Epub ahead of print
Periodico con rilevanza internazionale
pubmed
crossref
Aderisco
info:eu-repo/semantics/article
Long term complications in patients implanted with subcutaneous implantable defibrillators Real-world data from the Extended ELISIR experience / A. Gasperetti, M. Schiavone, M. Ziacchi, J. Vogler, A. Breitenstein, M. Laredo, P. Palmisano, D. Ricciardi, G. Mitacchione, P. Compagnucci, A. Bisignani, A. Angeletti, M. Casella, F. Picarelli, T. Fink, L. Kaiser, S. Hakmi, L. Calò, C. Pignalberi, L. Santini, C. Lavalle, E. Pisanò, I. Olivotto, C. Tondo, A. Curnis, A.D. Russo, N. Badenco, J. Steffel, C.J. Love, R. Tilz, G. Forleo, M. Biffi. - In: HEART RHYTHM. - ISSN 1547-5271. - (2021), pp. 1-9. [Epub ahead of print] [10.1016/j.hrthm.2021.07.008]
partially_open
Prodotti della ricerca::01 - Articolo su periodico
32
262
Article (author)
no
A. Gasperetti, M. Schiavone, M. Ziacchi, J. Vogler, A. Breitenstein, M. Laredo, P. Palmisano, D. Ricciardi, G. Mitacchione, P. Compagnucci, A. Bisigna...espandi
File in questo prodotto:
File Dimensione Formato  
long term.pdf

Open Access dal 15/07/2021

Tipologia: Post-print, accepted manuscript ecc. (versione accettata dall'editore)
Dimensione 8.53 MB
Formato Adobe PDF
8.53 MB Adobe PDF Visualizza/Apri
1-s2.0-S154752712101866X-main.pdf

accesso riservato

Tipologia: Publisher's version/PDF
Dimensione 1.58 MB
Formato Adobe PDF
1.58 MB Adobe PDF   Visualizza/Apri   Richiedi una copia
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/858923
Citazioni
  • ???jsp.display-item.citation.pmc??? 20
  • Scopus 45
  • ???jsp.display-item.citation.isi??? 38
  • OpenAlex ND
social impact