A microfluidic, label-free optical sensor for water pollutants, which is based on a packed micro-column of microspheres with refractive index similar to that of water, is presented. The perfluoropolyether microspheres are synthetized by membrane emulsification followed by UV irradiation. The microfluidic channel hosting the packed column is transparent when filled with pure water as a consequence of refractive index matching, whereas it scatters light in presence of compounds with lipophilic moieties that spontaneously adsorb on the fluorinated microspheres. The device is characterized by investigatingthe response to cationic and anionic surfactants. Both the signal growth rate and the recovery rate measured during washing with water depend on the type and concentration of the compounds. The cationic surfactants tested display a larger signal increase, linearly scaling with concentration. A limit of detection of 1muM is obtained in the current configuration. The water index-matched microspheres enable to access an additional analytical parameter, that is the propagation velocity of the scattering signal along the column. This parameter is also found to scale linearly with concentration, hence providing a complementary analytical tool sensitive to the adhesion kinetics.
A microfluidic column of water index-matched packed microspheres for label-free observation of water pollutants / R. Lanfranco, J. Saez, D. Abati, T. Carzaniga, F. Benito-Lopez, M. Buscaglia. - In: MIKROCHIMICA ACTA. - ISSN 0026-3672. - 188:4(2021 Mar 28). [10.1007/s00604-021-04804-2]
A microfluidic column of water index-matched packed microspheres for label-free observation of water pollutants
R. LanfrancoPrimo
;T. Carzaniga;M. Buscaglia
Ultimo
2021
Abstract
A microfluidic, label-free optical sensor for water pollutants, which is based on a packed micro-column of microspheres with refractive index similar to that of water, is presented. The perfluoropolyether microspheres are synthetized by membrane emulsification followed by UV irradiation. The microfluidic channel hosting the packed column is transparent when filled with pure water as a consequence of refractive index matching, whereas it scatters light in presence of compounds with lipophilic moieties that spontaneously adsorb on the fluorinated microspheres. The device is characterized by investigatingthe response to cationic and anionic surfactants. Both the signal growth rate and the recovery rate measured during washing with water depend on the type and concentration of the compounds. The cationic surfactants tested display a larger signal increase, linearly scaling with concentration. A limit of detection of 1muM is obtained in the current configuration. The water index-matched microspheres enable to access an additional analytical parameter, that is the propagation velocity of the scattering signal along the column. This parameter is also found to scale linearly with concentration, hence providing a complementary analytical tool sensitive to the adhesion kinetics.File | Dimensione | Formato | |
---|---|---|---|
Lanfranco2021_Article_AMicrofluidicColumnOfWaterInde.pdf
accesso riservato
Descrizione: Articolo pubblicato
Tipologia:
Publisher's version/PDF
Dimensione
1.68 MB
Formato
Adobe PDF
|
1.68 MB | Adobe PDF | Visualizza/Apri Richiedi una copia |
Microfluidic column of WIM packed beads_post_print.pdf
Open Access dal 29/03/2022
Descrizione: Articolo completo con supporting information
Tipologia:
Post-print, accepted manuscript ecc. (versione accettata dall'editore)
Dimensione
1.29 MB
Formato
Adobe PDF
|
1.29 MB | Adobe PDF | Visualizza/Apri |
Pubblicazioni consigliate
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.