Purpose: To further investigate the safety and performance of the Passeo-18 Lux drug-coated balloon (DCB) for the treatment of atherosclerotic infrainguinal disease under real-world conditions. Materials and Methods: BIOLUX P-III is an international, prospective, observational registry (ClinicalTrials.gov identifier NCT02276313) conducted at 41 centers in Europe, Asia, and Australia with follow-up visits at 6, 12, and 24 months. Of 700 patients (mean age 70.0±10.2 years; 439 men) with 863 lesions in the all-comers cohort, 330 (47.1%) patients had diabetes and 234 (37.7%) had chronic limb-threatening ischemia. The majority (79.3%) of lesions were in the femoropopliteal segment; of all lesions, 645 (74.9%) were calcified and 99 (11.5%) had in-stent restenosis (ISR). The mean lesion length was 84.7±73.3 mm. The primary clinical endpoint was major adverse events (MAEs) within 6 months, a composite of device- and procedure-related mortality through 30 days, major target limb amputation, and clinically-driven target lesion revascularization (TLR). The primary performance endpoint was clinically-driven TLR within 12 months. Results: At 6 and 12 months, freedom from MAEs was 94.0% and 89.5% in the all-comers cohort: 95.0% and 91.2% in the femoropopliteal group and 95.3% and 88.0% in the ISR subgroup, respectively. Freedom from clinically-driven TLR at 12 months was 93.1% in the all-comers cohort, 93.9% in the femoropopliteal lesions, and 89.4% for ISR lesions. All-cause mortality was 6.1% in the all-comers cohort: 5.9% in both the femoropopliteal and ISR subgroups. There were no device- or procedure-related deaths at up to 12 months. The Rutherford category improved in >80% of all subgroups at 12 months. Conclusion: In a real-world patient population, the safety and performance of the Passeo-18 Lux DCB for the treatment of atherosclerotic infrainguinal lesions are maintained, with good performance outcomes and low complication rates at 12 months.

Paclitaxel-Coated Balloon for the Treatment of Infrainguinal Disease: 12-Month Outcomes in the All-Comers Cohort of BIOLUX P-III Global Registry / G. Tepe, T. Zeller, M. Moscovic, J.-. Corpataux, J.K. Christensen, K. Keirse, G. Nano, H. Schroeder, C.A. Binkert, M. Brodmann. - In: JOURNAL OF ENDOVASCULAR THERAPY. - ISSN 1526-6028. - 27:2(2020 Apr), pp. 304-315. [10.1177/1526602819898804]

Paclitaxel-Coated Balloon for the Treatment of Infrainguinal Disease: 12-Month Outcomes in the All-Comers Cohort of BIOLUX P-III Global Registry

G. Nano;
2020

Abstract

Purpose: To further investigate the safety and performance of the Passeo-18 Lux drug-coated balloon (DCB) for the treatment of atherosclerotic infrainguinal disease under real-world conditions. Materials and Methods: BIOLUX P-III is an international, prospective, observational registry (ClinicalTrials.gov identifier NCT02276313) conducted at 41 centers in Europe, Asia, and Australia with follow-up visits at 6, 12, and 24 months. Of 700 patients (mean age 70.0±10.2 years; 439 men) with 863 lesions in the all-comers cohort, 330 (47.1%) patients had diabetes and 234 (37.7%) had chronic limb-threatening ischemia. The majority (79.3%) of lesions were in the femoropopliteal segment; of all lesions, 645 (74.9%) were calcified and 99 (11.5%) had in-stent restenosis (ISR). The mean lesion length was 84.7±73.3 mm. The primary clinical endpoint was major adverse events (MAEs) within 6 months, a composite of device- and procedure-related mortality through 30 days, major target limb amputation, and clinically-driven target lesion revascularization (TLR). The primary performance endpoint was clinically-driven TLR within 12 months. Results: At 6 and 12 months, freedom from MAEs was 94.0% and 89.5% in the all-comers cohort: 95.0% and 91.2% in the femoropopliteal group and 95.3% and 88.0% in the ISR subgroup, respectively. Freedom from clinically-driven TLR at 12 months was 93.1% in the all-comers cohort, 93.9% in the femoropopliteal lesions, and 89.4% for ISR lesions. All-cause mortality was 6.1% in the all-comers cohort: 5.9% in both the femoropopliteal and ISR subgroups. There were no device- or procedure-related deaths at up to 12 months. The Rutherford category improved in >80% of all subgroups at 12 months. Conclusion: In a real-world patient population, the safety and performance of the Passeo-18 Lux DCB for the treatment of atherosclerotic infrainguinal lesions are maintained, with good performance outcomes and low complication rates at 12 months.
chronic limb-threatening ischemia; critical limb ischemia; drug-coated balloon; drug-eluting balloon; femoropopliteal segment; paclitaxel; peripheral artery disease; superficial femoral artery
Settore MED/22 - Chirurgia Vascolare
apr-2020
gen-2020
Article (author)
File in questo prodotto:
File Dimensione Formato  
10.1177_1526602819898804.pdf

accesso aperto

Tipologia: Publisher's version/PDF
Dimensione 944.59 kB
Formato Adobe PDF
944.59 kB Adobe PDF Visualizza/Apri
15266028211014307.pdf

accesso aperto

Tipologia: Publisher's version/PDF
Dimensione 101.63 kB
Formato Adobe PDF
101.63 kB Adobe PDF Visualizza/Apri
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/736149
Citazioni
  • ???jsp.display-item.citation.pmc??? 1
  • Scopus 11
  • ???jsp.display-item.citation.isi??? 12
social impact