Background/aims To systematically compare the intermodality and inter-reader agreement for two blue-light confocal fundus autofluorescence (FAF) systems. Methods Thirty eyes (21 patients) with a diagnosis of geographic atrophy (GA) were enrolled. Eyes were imaged using two confocal blue-light FAF devices: (1) Spectralis device with a 488 nm excitation wavelength (488-FAF); (2) EIDON device with 450 nm excitation wavelength and the capability for 'colour' FAF imaging including both the individual red and green components of the emission spectrum. Furthermore, a third imaging modality (450-RF image) isolating and highlighting the red emission fluorescence component (REFC) was obtained and graded. Each image was graded by two readers to assess inter-reader variability and a single image for each modality was used to assess the intermodality variability. Results The 95% coefficient of repeatability (1.35 mm 2 for the 488-FAF-based grading, 8.13 mm 2 for the 450-FAF-based grading and 1.08 mm 2 for the 450-RF-based grading), the coefficient of variation (1.11 for 488-FAF, 2.05 for 450-FAF, 0.92 for 450-RF) and the intraclass correlation coefficient (0.994 for 488-FAF, 0.711 for 450-FAF, 0.997 for 450-RF) indicated that 450-FAF-based and 450-RF-based grading have the lowest and highest inter-reader agreements, respectively. The GA area was larger for 488-FAF images (median (IQR) 2.1 mm 2 (0.8-6.4 mm 2)) than for 450-FAF images (median (IQR) 1.0 mm 2 (0.3-4.3 mm 2); p<0.0001). There was no significant difference in lesion area measurement between 488-FAF-based and 450-RF-based grading (median (IQR) 2.6 mm 2 (0.8-6.8 mm 2); p=1.0). Conclusion The isolation of the REFC from the 450-FAF images allowed for a reproducible quantification of GA. This assessment had good comparability with that obtained with 488-FAF images.

Comparison of short-wavelength blue-light autofluorescence and conventional blue-light autofluorescence in geographic atrophy / E. Borrelli, M.G. Nittala, N.S. Abdelfattah, J. Lei, A.H. Hariri, Y. Shi, W. Fan, M. Cozzi, V. Sarao, P. Lanzetta, G. Staurenghi, S.R. Sadda. - In: BRITISH JOURNAL OF OPHTHALMOLOGY. - ISSN 0007-1161. - 103:5(2019), pp. 610-616. [10.1136/bjophthalmol-2018-311849]

Comparison of short-wavelength blue-light autofluorescence and conventional blue-light autofluorescence in geographic atrophy

M. Cozzi;G. Staurenghi;
2019

Abstract

Background/aims To systematically compare the intermodality and inter-reader agreement for two blue-light confocal fundus autofluorescence (FAF) systems. Methods Thirty eyes (21 patients) with a diagnosis of geographic atrophy (GA) were enrolled. Eyes were imaged using two confocal blue-light FAF devices: (1) Spectralis device with a 488 nm excitation wavelength (488-FAF); (2) EIDON device with 450 nm excitation wavelength and the capability for 'colour' FAF imaging including both the individual red and green components of the emission spectrum. Furthermore, a third imaging modality (450-RF image) isolating and highlighting the red emission fluorescence component (REFC) was obtained and graded. Each image was graded by two readers to assess inter-reader variability and a single image for each modality was used to assess the intermodality variability. Results The 95% coefficient of repeatability (1.35 mm 2 for the 488-FAF-based grading, 8.13 mm 2 for the 450-FAF-based grading and 1.08 mm 2 for the 450-RF-based grading), the coefficient of variation (1.11 for 488-FAF, 2.05 for 450-FAF, 0.92 for 450-RF) and the intraclass correlation coefficient (0.994 for 488-FAF, 0.711 for 450-FAF, 0.997 for 450-RF) indicated that 450-FAF-based and 450-RF-based grading have the lowest and highest inter-reader agreements, respectively. The GA area was larger for 488-FAF images (median (IQR) 2.1 mm 2 (0.8-6.4 mm 2)) than for 450-FAF images (median (IQR) 1.0 mm 2 (0.3-4.3 mm 2); p<0.0001). There was no significant difference in lesion area measurement between 488-FAF-based and 450-RF-based grading (median (IQR) 2.6 mm 2 (0.8-6.8 mm 2); p=1.0). Conclusion The isolation of the REFC from the 450-FAF images allowed for a reproducible quantification of GA. This assessment had good comparability with that obtained with 488-FAF images.
Age-related macular degeneration; Autofluorescence; Geographic atrophy; Green fluorophores; Optical coherence tomography
Settore MED/30 - Malattie Apparato Visivo
2019
Article (author)
File in questo prodotto:
File Dimensione Formato  
610.full.pdf

accesso aperto

Tipologia: Publisher's version/PDF
Dimensione 1.87 MB
Formato Adobe PDF
1.87 MB Adobe PDF Visualizza/Apri
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/704335
Citazioni
  • ???jsp.display-item.citation.pmc??? 6
  • Scopus 24
  • ???jsp.display-item.citation.isi??? 22
social impact