OBJECTIVES: The main reason for aortic repair failures is recurrent annular dilatation. The fibrous portion of left ventricular outflow tract dilates. A novel device was designed to tackle this problem. METHODS: The device consists of an internal ring applied at the aortic annulus plus an external flexible band at the level of the aortic root. The internal ring has a semi-rigid portion (40%, placed at ventriculo-arterial junction) and a flexible portion to allow it to conform along the curves of the non-coronary/right coronary leaflet and right coronary/left coronary leaflet commissures. The external band acts as a reinforcement to the internal ring. A pulsatile mock loop capable of housing porcine aortic valve was used. Working conditions were 60 bpm of heart rate, 75 of stroke volumes and 120-80mmHg of simulated pressure. Mean gradient, effective orifice area, annular diameter, coaptation height and length were recorded on 11 aortic root units (ARUs). High-speed video and standard echocardiographic images were also recorded. All data were acquired in the following conditions: (i) basal (untreated ARU); (ii) pathological condition (left coronary/non-coronary triangle was dilated by suturing an aortic patch); and (iii) ARU treated with the device. RESULTS: Gradients and effective orifice area were respectively 0.9 ± 0.64mmHg and 3.1 ± 0.7cm2 (pathological) and 3.7 ± 1.1mmHg and 1.5 ± 0.2cm2 (treated, P < 0.05). Left coronary/non-coronary diameter decreased from 2.4 ± 0.2 cm (pathological) to 2.0 ± 0.2 (treated, P < 0.05). Coaptation length and height were fully restored to basal values following treatment. Visual inspection showed proper dynamics of the leaflet, confirmed by high-speed video and echocardiography. CONCLUSIONS: The device allowed for restoring physiologic-like coaptation in the experimental model, without inducing clinically relevant worsening of the haemodynamics of the treated ARU.

A novel system for the treatment of aortic annular dilation : an ex vivo investigation / P. Shah, C. Romagnoni, M. Jaworek, F. Lucherini, M. Contino, A. Menkis, G. Gelpi, G.B. Fiore, C. Antona, R. Vismara. - In: EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY. - ISSN 1873-734X. - 52:6(2017 Dec), pp. 1090-1097. [10.1093/ejcts/ezx203]

A novel system for the treatment of aortic annular dilation : an ex vivo investigation

C. Romagnoni;M. Contino;C. Antona;
2017

Abstract

OBJECTIVES: The main reason for aortic repair failures is recurrent annular dilatation. The fibrous portion of left ventricular outflow tract dilates. A novel device was designed to tackle this problem. METHODS: The device consists of an internal ring applied at the aortic annulus plus an external flexible band at the level of the aortic root. The internal ring has a semi-rigid portion (40%, placed at ventriculo-arterial junction) and a flexible portion to allow it to conform along the curves of the non-coronary/right coronary leaflet and right coronary/left coronary leaflet commissures. The external band acts as a reinforcement to the internal ring. A pulsatile mock loop capable of housing porcine aortic valve was used. Working conditions were 60 bpm of heart rate, 75 of stroke volumes and 120-80mmHg of simulated pressure. Mean gradient, effective orifice area, annular diameter, coaptation height and length were recorded on 11 aortic root units (ARUs). High-speed video and standard echocardiographic images were also recorded. All data were acquired in the following conditions: (i) basal (untreated ARU); (ii) pathological condition (left coronary/non-coronary triangle was dilated by suturing an aortic patch); and (iii) ARU treated with the device. RESULTS: Gradients and effective orifice area were respectively 0.9 ± 0.64mmHg and 3.1 ± 0.7cm2 (pathological) and 3.7 ± 1.1mmHg and 1.5 ± 0.2cm2 (treated, P < 0.05). Left coronary/non-coronary diameter decreased from 2.4 ± 0.2 cm (pathological) to 2.0 ± 0.2 (treated, P < 0.05). Coaptation length and height were fully restored to basal values following treatment. Visual inspection showed proper dynamics of the leaflet, confirmed by high-speed video and echocardiography. CONCLUSIONS: The device allowed for restoring physiologic-like coaptation in the experimental model, without inducing clinically relevant worsening of the haemodynamics of the treated ARU.
Annuloplasty ring; Aortic valve; Aortic valve repair; Animals; Aortic Valve; Cardiac Valve Annuloplasty; Dilatation, Pathologic; Disease Models, Animal; Echocardiography; Heart Valve Diseases; Swine; Treatment Outcome
Settore MED/23 - Chirurgia Cardiaca
dic-2017
Article (author)
File in questo prodotto:
File Dimensione Formato  
ezx203.pdf

accesso aperto

Tipologia: Publisher's version/PDF
Dimensione 713.98 kB
Formato Adobe PDF
713.98 kB Adobe PDF Visualizza/Apri
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/665337
Citazioni
  • ???jsp.display-item.citation.pmc??? 2
  • Scopus 3
  • ???jsp.display-item.citation.isi??? 3
social impact