We design and manufacture a silicone model of the human aorta, able to mimic both the geometrical and the mechanical properties of physiological individuals, with a specific focus on reproducing the compliance. In fact, while the models available in the literature exhibit an unrealistic compliant behavior, though they are detailed from the geometrical viewpoint, here the goal is to provide an accurate compliant tool for in vitro testing the devices that interface with the vascular system. A parametric design of the aortic model is obtained based on the available literature data, and the model is manufactured with a specific silicone mixture using rapid prototyping and molding techniques. The manufactured prototype has been tested by means of computed tomography scans for evaluating the matching of the mechanical properties with the desired ones. Results show a high degree of adherence between the imposed and the measured compliance values for each main aortic section. Thus, our work proves the feasibility of the approach, and the possibility to manufacture compliant models that reproduce the mechanical behavior of the aorta for in vitro studies.

A compliant aortic model for in vitro simulations : design and manufacturing process / S. Marconi, E. Lanzarone, G.H.W. van Bogerijen, M. Conti, F. Secchi, S. Trimarchi, F. Auricchio. - In: MEDICAL ENGINEERING & PHYSICS. - ISSN 1350-4533. - 59(2018 Sep), pp. 21-29. [10.1016/j.medengphy.2018.04.022]

A compliant aortic model for in vitro simulations : design and manufacturing process

F. Secchi;S. Trimarchi
Penultimo
;
2018

Abstract

We design and manufacture a silicone model of the human aorta, able to mimic both the geometrical and the mechanical properties of physiological individuals, with a specific focus on reproducing the compliance. In fact, while the models available in the literature exhibit an unrealistic compliant behavior, though they are detailed from the geometrical viewpoint, here the goal is to provide an accurate compliant tool for in vitro testing the devices that interface with the vascular system. A parametric design of the aortic model is obtained based on the available literature data, and the model is manufactured with a specific silicone mixture using rapid prototyping and molding techniques. The manufactured prototype has been tested by means of computed tomography scans for evaluating the matching of the mechanical properties with the desired ones. Results show a high degree of adherence between the imposed and the measured compliance values for each main aortic section. Thus, our work proves the feasibility of the approach, and the possibility to manufacture compliant models that reproduce the mechanical behavior of the aorta for in vitro studies.
Aortic model; Compliance reproduction; In vitro simulation; Molding; Parametric design; Rapid prototyping; Aorta; Compliance; Materials Testing; Silicones; Computer-Aided Design; Models, Anatomic
Settore MED/36 - Diagnostica per Immagini e Radioterapia
Settore ING-IND/22 - Scienza e Tecnologia dei Materiali
set-2018
Article (author)
File in questo prodotto:
File Dimensione Formato  
main.pdf

accesso riservato

Tipologia: Publisher's version/PDF
Dimensione 1.62 MB
Formato Adobe PDF
1.62 MB Adobe PDF   Visualizza/Apri   Richiedi una copia
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/660601
Citazioni
  • ???jsp.display-item.citation.pmc??? 2
  • Scopus 18
  • ???jsp.display-item.citation.isi??? 13
social impact