The fast neutrons produced by a calibrated 241Am-Be source were detected by two different Cs2LiYCl6:Ce (CLYC) scintillator detectors. The two cylindrical crystals (1′′× 1′′ in size) were enriched with more than 99% of 7Li (C7LYC) and with about 95% of 6Li (C6LYC), respectively. Both crystals can detect fast neutrons whereas only C6LYC can also detect thermal neutrons, due to the presence of 6Li. The measurement was performed at the L.A.S.A . Laboratory of INFN and University of Milano (Italy). To identify the neutron events, the Pulse-Shape-Discrimination technique was used. The values of 1.41 ± 0.16 for C6LYC and 1.16 ± 0.21 for C7LYC for the detection efficiency of the 241Am-Be emitted neutrons, with energy up to 10 MeV, were deduced.
Fast neutron detection efficiency of 6Li and 7Li enriched CLYC scintillators using an Am-Be source / N. Blasi, S. Brambilla, F. Camera, S. Ceruti, A. Giaz, L. Gini, F. Groppi, S. Manenti, A. Mentana, B. Million, S. Riboldi. - In: JOURNAL OF INSTRUMENTATION. - ISSN 1748-0221. - 13:11(2018 Nov). [10.1088/1748-0221/13/11/P11010]
Fast neutron detection efficiency of 6Li and 7Li enriched CLYC scintillators using an Am-Be source
F. Camera
;S. Ceruti;A. Giaz;F. Groppi;S. Manenti;A. Mentana;S. RiboldiUltimo
2018
Abstract
The fast neutrons produced by a calibrated 241Am-Be source were detected by two different Cs2LiYCl6:Ce (CLYC) scintillator detectors. The two cylindrical crystals (1′′× 1′′ in size) were enriched with more than 99% of 7Li (C7LYC) and with about 95% of 6Li (C6LYC), respectively. Both crystals can detect fast neutrons whereas only C6LYC can also detect thermal neutrons, due to the presence of 6Li. The measurement was performed at the L.A.S.A . Laboratory of INFN and University of Milano (Italy). To identify the neutron events, the Pulse-Shape-Discrimination technique was used. The values of 1.41 ± 0.16 for C6LYC and 1.16 ± 0.21 for C7LYC for the detection efficiency of the 241Am-Be emitted neutrons, with energy up to 10 MeV, were deduced.File | Dimensione | Formato | |
---|---|---|---|
Blasi_2018_J._Inst._13_P11010.pdf
accesso riservato
Tipologia:
Publisher's version/PDF
Dimensione
1.15 MB
Formato
Adobe PDF
|
1.15 MB | Adobe PDF | Visualizza/Apri Richiedi una copia |
Pubblicazioni consigliate
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.