AIM OF THE STUDY: To compare the diagnostic performance of a commercially available computer-aided diagnosis (CAD) system for thyroid ultrasound (US) with that of a non-computer-aided radiologist in the characterization of low-to-high suspicion thyroid nodules. METHODS: This retrospective study included a consecutive series of adult patients referred for US-guided fine-needle aspiration biopsy (FNAB) of a thyroid nodule. All patients were eligible for thyroid nodule FNAB according to the current international guidelines. An interventional radiologist experienced in thyroid imaging acquired the US images subsequently used for post-processing, performed FNAB and provided the US features of each nodule. A radiology resident and an endocrinology resident in consensus performed post-processing using the CAD system to assess the same nodule characteristics. The diagnostic performance and agreement of US features between the CAD system and the radiologist were compared. RESULTS: Sixty-two patients (50 F; age 60 ± 12 years) were enrolled: 77.4% (48/62) of thyroid nodules were benign, 22.6% (14/62) were undetermined to malignant and required follow-up or surgery. Interobserver agreement between the CAD system and the radiologist was substantial for orientation (K = 0.69), fair for composition (K = 0.36), echogenicity (K = 0.36), K-TIRADS (K = 0.29), and slight for margins (K = 0.03). The radiologist demonstrated a significantly higher sensitivity than the CAD system (78.6% vs. 21.4%; P = 0.008), while there was no statistical difference in specificity (66.7% vs. 81.3%; P = 0.065). CONCLUSION: This CAD system is less sensitive than an experienced radiologist and showed slight-to-substantial agreement with the radiologist for the characterization of thyroid nodules. Although it is an innovative tool with good potential, additional efforts are needed to improve its diagnostic performance.

A computer-aided diagnosis system for the assessment and characterization of low-to-high suspicion thyroid nodules on ultrasound / S. Gitto, G. Grassi, C. De Angelis, C.G. Monaco, S. Sdao, F. Sardanelli, L.M. Sconfienza, G. Mauri. - In: LA RADIOLOGIA MEDICA. - ISSN 0033-8362. - 124:2(2019 Feb), pp. 118-125. [10.1007/s11547-018-0942-z]

A computer-aided diagnosis system for the assessment and characterization of low-to-high suspicion thyroid nodules on ultrasound

S. Gitto
;
G. Grassi;C. De Angelis;C.G. Monaco;S. Sdao;F. Sardanelli;L.M. Sconfienza;G. Mauri
2019

Abstract

AIM OF THE STUDY: To compare the diagnostic performance of a commercially available computer-aided diagnosis (CAD) system for thyroid ultrasound (US) with that of a non-computer-aided radiologist in the characterization of low-to-high suspicion thyroid nodules. METHODS: This retrospective study included a consecutive series of adult patients referred for US-guided fine-needle aspiration biopsy (FNAB) of a thyroid nodule. All patients were eligible for thyroid nodule FNAB according to the current international guidelines. An interventional radiologist experienced in thyroid imaging acquired the US images subsequently used for post-processing, performed FNAB and provided the US features of each nodule. A radiology resident and an endocrinology resident in consensus performed post-processing using the CAD system to assess the same nodule characteristics. The diagnostic performance and agreement of US features between the CAD system and the radiologist were compared. RESULTS: Sixty-two patients (50 F; age 60 ± 12 years) were enrolled: 77.4% (48/62) of thyroid nodules were benign, 22.6% (14/62) were undetermined to malignant and required follow-up or surgery. Interobserver agreement between the CAD system and the radiologist was substantial for orientation (K = 0.69), fair for composition (K = 0.36), echogenicity (K = 0.36), K-TIRADS (K = 0.29), and slight for margins (K = 0.03). The radiologist demonstrated a significantly higher sensitivity than the CAD system (78.6% vs. 21.4%; P = 0.008), while there was no statistical difference in specificity (66.7% vs. 81.3%; P = 0.065). CONCLUSION: This CAD system is less sensitive than an experienced radiologist and showed slight-to-substantial agreement with the radiologist for the characterization of thyroid nodules. Although it is an innovative tool with good potential, additional efforts are needed to improve its diagnostic performance.
Computer-aided diagnosis; Nodule; Thyroid; Ultrasound
Settore MED/36 - Diagnostica per Immagini e Radioterapia
feb-2019
Article (author)
File in questo prodotto:
File Dimensione Formato  
10.1007@s11547-018-0942-z.pdf

accesso riservato

Tipologia: Publisher's version/PDF
Dimensione 1.11 MB
Formato Adobe PDF
1.11 MB Adobe PDF   Visualizza/Apri   Richiedi una copia
Gitto2019_Article_AComputer-aidedDiagnosisSystem.pdf

accesso riservato

Tipologia: Publisher's version/PDF
Dimensione 1.11 MB
Formato Adobe PDF
1.11 MB Adobe PDF   Visualizza/Apri   Richiedi una copia
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/591037
Citazioni
  • ???jsp.display-item.citation.pmc??? 25
  • Scopus 43
  • ???jsp.display-item.citation.isi??? 37
social impact