BACKGROUND: Early results of the Micra Investigational Device Exemption (IDE) study and Micra Post-Approval Registry (PAR) demonstrated excellent safety and efficacy performance; however, intermediate-term results across a large patient population in the real-world setting have not been evaluated. OBJECTIVES: We report updated performance of the Micra transcatheter pacemaker from a worldwide PAR and compare it with the IDE study as well as a transvenous historical control. METHODS: The safety objective of the analysis was system- or procedure-related major complications through 12 months postimplantation. We compared the major complication rate with that of the 726 patients from the IDE and with a reference data set of 2667 patients with transvenous pacemakers by using a Fine-Gray competing risk model. RESULTS: The Micra device was successfully implanted in 1801 of 1817 patients (99.1%). The mean follow-up period was 6.8 ± 6.9 months. Through 12 months, the major complication rate was 2.7% (95% confidence interval [CI] 2.0%-3.7%). The risk of major complications for Micra PAR patients was 63% lower than that for patients with transvenous pacemakers through 12 months postimplantation (hazard ratio 0.37; 95% CI 0.27-0.52; P < .001). The major complication rate trended lower in the PAR than in the IDE study (hazard ratio 0.71; 95% CI 0.44-1.1; P = .160), driven by the lower pericardial effusion rate in the PAR. There were 3 cases of infection associated with the procedure, but none required device removal and there were no battery or telemetry issues. Pacing thresholds were low and stable through 12 months postimplantation. CONCLUSION: Performance of the Micra transcatheter pacemaker in international clinical practice remains consistent with previously reported data. Major complications were infrequent and occurred 63% less often compared to transvenous systems.

Updated Performance of the Micra Transcatheter Pacemaker in the Real-World Setting : a Comparison to the Investigational Study and a Transvenous Historical Control / M.F. El-Chami, F. Al-Samadi, N. Clementy, C. Garweg, J.L. Martinez-Sande, J.P. Piccini, S. Iacopino, M. Lloyd, X. Viñolas Prat, M.D. Jacobsen, P. Ritter, J.B. Johansen, C. Tondo, F. Liu, D.H. Fagan, A.K. Eakley, P.R. Roberts. - In: HEART RHYTHM. - ISSN 1547-5271. - 15:12(2018 Dec), pp. 1800-1807.

Updated Performance of the Micra Transcatheter Pacemaker in the Real-World Setting : a Comparison to the Investigational Study and a Transvenous Historical Control

C. Tondo;
2018

Abstract

BACKGROUND: Early results of the Micra Investigational Device Exemption (IDE) study and Micra Post-Approval Registry (PAR) demonstrated excellent safety and efficacy performance; however, intermediate-term results across a large patient population in the real-world setting have not been evaluated. OBJECTIVES: We report updated performance of the Micra transcatheter pacemaker from a worldwide PAR and compare it with the IDE study as well as a transvenous historical control. METHODS: The safety objective of the analysis was system- or procedure-related major complications through 12 months postimplantation. We compared the major complication rate with that of the 726 patients from the IDE and with a reference data set of 2667 patients with transvenous pacemakers by using a Fine-Gray competing risk model. RESULTS: The Micra device was successfully implanted in 1801 of 1817 patients (99.1%). The mean follow-up period was 6.8 ± 6.9 months. Through 12 months, the major complication rate was 2.7% (95% confidence interval [CI] 2.0%-3.7%). The risk of major complications for Micra PAR patients was 63% lower than that for patients with transvenous pacemakers through 12 months postimplantation (hazard ratio 0.37; 95% CI 0.27-0.52; P < .001). The major complication rate trended lower in the PAR than in the IDE study (hazard ratio 0.71; 95% CI 0.44-1.1; P = .160), driven by the lower pericardial effusion rate in the PAR. There were 3 cases of infection associated with the procedure, but none required device removal and there were no battery or telemetry issues. Pacing thresholds were low and stable through 12 months postimplantation. CONCLUSION: Performance of the Micra transcatheter pacemaker in international clinical practice remains consistent with previously reported data. Major complications were infrequent and occurred 63% less often compared to transvenous systems.
leadless pacing; transvenous pacemaker; real-world performance; transcatheter pacemaker; updated results
Settore MED/11 - Malattie dell'Apparato Cardiovascolare
dic-2018
Article (author)
File in questo prodotto:
File Dimensione Formato  
1-s2.0-S1547527118307926-main.pdf

accesso aperto

Tipologia: Publisher's version/PDF
Dimensione 562.22 kB
Formato Adobe PDF
562.22 kB Adobe PDF Visualizza/Apri
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/585808
Citazioni
  • ???jsp.display-item.citation.pmc??? 100
  • Scopus 263
  • ???jsp.display-item.citation.isi??? 215
social impact