The Investigations of this article focus on the response of an Al2O3:C radioluminescence (RL) prototype for medical dosimetry in a 6 MV photon beam. The prototype can be configured using two types of detectors coupled to fiber-optic cables - single crystal (1 x 1 x 2 mm(3)) and droplets (in two grain sizes, 38 and 4 mu m, molded in r =0.5 mm,1= 200 mu m). By using the appropriate filters in addition to time gating it is possible to remove disturbance present during irradiation: the stem effect. Pre -irradiation of the dosimeters to a dose of 300 Gy made the memory effects in Al2O3:C negligible, so as to not impair the dosimetric properties of the system. The key findings are that the system is suitable for small field beam dosimetry, while giving overall good dose response in other features (i.e., beam profile, dose rate - FF and FFF modes). The results show that our prototype can be used for real time dose rate assessment in medical photon dosimetry without many correction factors. The 41 mu m RL measurement results are in excellent agreement (i.e. below 1%) with the dose delivered according to standard beam data.

Radioluminescence results from an Al2O3:C fiber prototype: 6 MV medical beam / L.F. Nascimento, I. Veronese, G. Loi, E. Mones, F. Vanhavere, D. Verellen. - In: SENSORS AND ACTUATORS. A, PHYSICAL. - ISSN 0924-4247. - 274(2018 May 01), pp. 1-9. [10.1016/j.sna.2018.03.007]

Radioluminescence results from an Al2O3:C fiber prototype: 6 MV medical beam

I. Veronese;
2018

Abstract

The Investigations of this article focus on the response of an Al2O3:C radioluminescence (RL) prototype for medical dosimetry in a 6 MV photon beam. The prototype can be configured using two types of detectors coupled to fiber-optic cables - single crystal (1 x 1 x 2 mm(3)) and droplets (in two grain sizes, 38 and 4 mu m, molded in r =0.5 mm,1= 200 mu m). By using the appropriate filters in addition to time gating it is possible to remove disturbance present during irradiation: the stem effect. Pre -irradiation of the dosimeters to a dose of 300 Gy made the memory effects in Al2O3:C negligible, so as to not impair the dosimetric properties of the system. The key findings are that the system is suitable for small field beam dosimetry, while giving overall good dose response in other features (i.e., beam profile, dose rate - FF and FFF modes). The results show that our prototype can be used for real time dose rate assessment in medical photon dosimetry without many correction factors. The 41 mu m RL measurement results are in excellent agreement (i.e. below 1%) with the dose delivered according to standard beam data.
radioluminescence; radiotherapy; real time dosimetry; droplets; small field
Settore FIS/07 - Fisica Applicata(Beni Culturali, Ambientali, Biol.e Medicin)
1-mag-2018
Article (author)
File in questo prodotto:
File Dimensione Formato  
SNAA_2018_accepted.pdf

accesso aperto

Tipologia: Pre-print (manoscritto inviato all'editore)
Dimensione 811.1 kB
Formato Adobe PDF
811.1 kB Adobe PDF Visualizza/Apri
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/567296
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 13
  • ???jsp.display-item.citation.isi??? 13
social impact