We report the case of a 50 year-old woman referred for extraction of an infected biventricular implantable cardioverter defibrillator (ICD) who underwent device and leads extraction. Due to adhesions at the level of the subclavian–innominate vein angle, manual traction with locking stylet was ineffective. A new mechanical sheath provided with a stainless steel bladed tip, the Evolution™ Mechanical Dilator Sheath (Cook Medical), was successfully used to complete lead extraction. As far as we know, this is the first report on the use of this new tool; its safety should be ascertained by increasing the number of cases but its simple use and its low cost make the Evolution™ Mechanical Dilator Sheath a new interesting tool that can be added to the instruments available for lead extraction.

Lead extraction : a new effective tool to overcome fibrous binding sites / D.R. Antonio, B. Roberto, P. Gemma, C. Michela, F. Antonio, C. Antonella, B. Caterina, F. Giovanni B., C. Tondo, Z. Paolo, B. Fulvio. - In: JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY. - ISSN 1383-875X. - 24:2(2009 Mar), pp. 147-150.

Lead extraction : a new effective tool to overcome fibrous binding sites

C. Tondo;
2009

Abstract

We report the case of a 50 year-old woman referred for extraction of an infected biventricular implantable cardioverter defibrillator (ICD) who underwent device and leads extraction. Due to adhesions at the level of the subclavian–innominate vein angle, manual traction with locking stylet was ineffective. A new mechanical sheath provided with a stainless steel bladed tip, the Evolution™ Mechanical Dilator Sheath (Cook Medical), was successfully used to complete lead extraction. As far as we know, this is the first report on the use of this new tool; its safety should be ascertained by increasing the number of cases but its simple use and its low cost make the Evolution™ Mechanical Dilator Sheath a new interesting tool that can be added to the instruments available for lead extraction.
implantable cardioverter defibrillator; infection; lead extraction; mechanical sheath; adult; defibrillators, implantable; device removal; electrodes, implanted; equipment design; equipment failure analysis; female; fibrosis; heart ventricles; humans; cardiology and cardiovascular medicine; physiology (medical)
Settore MED/11 - Malattie dell'Apparato Cardiovascolare
mar-2009
Article (author)
File in questo prodotto:
File Dimensione Formato  
Lead extraction.pdf

accesso riservato

Tipologia: Publisher's version/PDF
Dimensione 185.1 kB
Formato Adobe PDF
185.1 kB Adobe PDF   Visualizza/Apri   Richiedi una copia
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/540936
Citazioni
  • ???jsp.display-item.citation.pmc??? 3
  • Scopus 15
  • ???jsp.display-item.citation.isi??? 13
social impact