OBJECTIVES: Thoracic endovascular aortic repair (TEVAR) has been shown to lead to increased aortic stiffness. The aim of this study was to investigate the effect of stent graft type and stent graft length on aortic stiffness in a controlled, experimental setting. METHODS: Twenty porcine thoracic aortas were connected to a pulsatile mock loop system. Intraluminal pressure was recorded at two sites in order to measure pulse wave velocity (PWV) for each aorta: before stent graft deployment (t1); after deployment of a 100-mm long stent graft (t2); and after distal extension through deployment of a second 100-mm long stent graft (t3). Four different types of stent grafts (Conformable Gore® TAG® Device, Bolton Relay® Device, Cook Zenith Alpha™, and Medtronic Valiant®) were evaluated. RESULTS: For the total cohort of 20 aortas, PWV increased by a mean 0.6 m/s or 8.9% of baseline PWV after deployment of a 100-mm proximal stent graft (P<0.001), and by a mean 1.4 m/s or 23.0% of baseline PWV after distal extension of the stent graft (P<0.001). Univariable regression analysis showed a significant correlation between aortic PWV and extent of stent graft coverage, (P<0.001), but no significant effect of baseline aortic length, baseline aortic PWV, or stent graft type on the percentual increase in PWV at t2 or at t3. CONCLUSIONS: In this experimental set-up, aortic stiffness increased significantly after stent graft deployment with each of the four types of stent graft, with the increase in aortic stiffness depending on the extent of stent graft coverage.

Changes in aortic pulse wave velocity of four thoracic aortic stent grafts in an ex vivo porcine model / H..W..L. De Beaufort, M. Coda, M. Conti, T..M..J. Van Bakel, F..J..H. Nauta, E. Lanzarone, F..L. Moll, J..A. Van Herwaarden, F. Auricchio, S. Trimarchi. - In: PLOS ONE. - ISSN 1932-6203. - 12:10(2017 Oct 05), pp. e0186080.1-e0186080.7.

Changes in aortic pulse wave velocity of four thoracic aortic stent grafts in an ex vivo porcine model

F. Auricchio;S. Trimarchi
2017

Abstract

OBJECTIVES: Thoracic endovascular aortic repair (TEVAR) has been shown to lead to increased aortic stiffness. The aim of this study was to investigate the effect of stent graft type and stent graft length on aortic stiffness in a controlled, experimental setting. METHODS: Twenty porcine thoracic aortas were connected to a pulsatile mock loop system. Intraluminal pressure was recorded at two sites in order to measure pulse wave velocity (PWV) for each aorta: before stent graft deployment (t1); after deployment of a 100-mm long stent graft (t2); and after distal extension through deployment of a second 100-mm long stent graft (t3). Four different types of stent grafts (Conformable Gore® TAG® Device, Bolton Relay® Device, Cook Zenith Alpha™, and Medtronic Valiant®) were evaluated. RESULTS: For the total cohort of 20 aortas, PWV increased by a mean 0.6 m/s or 8.9% of baseline PWV after deployment of a 100-mm proximal stent graft (P<0.001), and by a mean 1.4 m/s or 23.0% of baseline PWV after distal extension of the stent graft (P<0.001). Univariable regression analysis showed a significant correlation between aortic PWV and extent of stent graft coverage, (P<0.001), but no significant effect of baseline aortic length, baseline aortic PWV, or stent graft type on the percentual increase in PWV at t2 or at t3. CONCLUSIONS: In this experimental set-up, aortic stiffness increased significantly after stent graft deployment with each of the four types of stent graft, with the increase in aortic stiffness depending on the extent of stent graft coverage.
Biochemistry, Genetics and Molecular Biology (all); Agricultural and Biological Sciences (all)
Settore MED/22 - Chirurgia Vascolare
5-ott-2017
Article (author)
File in questo prodotto:
File Dimensione Formato  
2017-PlosOne deBeaufort_Trimarchi ex_vivo_PWW_TEVAR.pdf

accesso aperto

Descrizione: Published article
Tipologia: Publisher's version/PDF
Dimensione 3.53 MB
Formato Adobe PDF
3.53 MB Adobe PDF Visualizza/Apri