The measurement of plasma glucose (PG) plays a central role in recognizing disturbances in carbohydrate metabolism, with established decision limits that are globally accepted. This requires that PG results are reliable and unequivocally valid no matter where they are obtained. To control the pre-analytical variability of PG and preventin vitro glycolysis, the use of citrate as rapidly effective glycolysis inhibitor has been proposed. However, the commercial availability of several tubes with studies showing different performance has created confusion among users. Moreover, and more importantly, studies have shown that tubes promptly inhibiting glycolysis give PG results that are significantly higher than tubes containing sodium fluoride only, used in the majority of studies generating the current PG cut-points, with a different clinical classification of subjects. From the analytical point of view, to be equivalent among different measuring systems, PG results should be traceable to a recognized higher-order reference via the implementation of an unbroken metrological hierarchy. In doing this, it is important that manufacturers of measuring systems consider the uncertainty accumulated through the different steps of the selected traceability chain. In particular, PG results should fulfil analytical performance specifications defined to fit the intended clinical application. Since PG has tight homeostatic control, its biological variability may be used to define these limits. Alternatively, given the central diagnostic role of the analyte, an outcome model showing the impact of analytical performance of test on clinical classifications of subjects can be used. Using these specifications, performance assessment studies employing commutable control materials with values assigned by reference procedure have shown that the quality of PG measurements is often far from desirable and that problems are exacerbated using point-of-care devices.

Pre-analytical and analytical aspects affecting clinical reliability of plasma glucose results / S. Pasqualetti, F. Braga, M. Panteghini. - In: CLINICAL BIOCHEMISTRY. - ISSN 0009-9120. - 50:10-11(2017), pp. 587-594. [10.1016/j.clinbiochem.2017.03.009]

Pre-analytical and analytical aspects affecting clinical reliability of plasma glucose results

F. Braga;M. Panteghini
2017

Abstract

The measurement of plasma glucose (PG) plays a central role in recognizing disturbances in carbohydrate metabolism, with established decision limits that are globally accepted. This requires that PG results are reliable and unequivocally valid no matter where they are obtained. To control the pre-analytical variability of PG and preventin vitro glycolysis, the use of citrate as rapidly effective glycolysis inhibitor has been proposed. However, the commercial availability of several tubes with studies showing different performance has created confusion among users. Moreover, and more importantly, studies have shown that tubes promptly inhibiting glycolysis give PG results that are significantly higher than tubes containing sodium fluoride only, used in the majority of studies generating the current PG cut-points, with a different clinical classification of subjects. From the analytical point of view, to be equivalent among different measuring systems, PG results should be traceable to a recognized higher-order reference via the implementation of an unbroken metrological hierarchy. In doing this, it is important that manufacturers of measuring systems consider the uncertainty accumulated through the different steps of the selected traceability chain. In particular, PG results should fulfil analytical performance specifications defined to fit the intended clinical application. Since PG has tight homeostatic control, its biological variability may be used to define these limits. Alternatively, given the central diagnostic role of the analyte, an outcome model showing the impact of analytical performance of test on clinical classifications of subjects can be used. Using these specifications, performance assessment studies employing commutable control materials with values assigned by reference procedure have shown that the quality of PG measurements is often far from desirable and that problems are exacerbated using point-of-care devices.
Glucose; Analytical performance specifications; Standardization; Uncertainty; EQAS
Settore BIO/12 - Biochimica Clinica e Biologia Molecolare Clinica
2017
Centro per la riferibilità Metrologica in Medicina di Laboratorio (CIRME)
Article (author)
File in questo prodotto:
File Dimensione Formato  
Pre-analytical and analytical aspects affecting clinical reliability of plasma glucose.pdf

accesso riservato

Tipologia: Publisher's version/PDF
Dimensione 378.37 kB
Formato Adobe PDF
378.37 kB Adobe PDF   Visualizza/Apri   Richiedi una copia
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/504578
Citazioni
  • ???jsp.display-item.citation.pmc??? 7
  • Scopus 27
  • ???jsp.display-item.citation.isi??? 22
social impact