Aims: Insulin pump failure and/or malfunction requiring replacement have not been thoroughly investigated. This study evaluated pump replacement in children and adolescents with Type 1 diabetes using insulin pump therapy. Methods: Data were collected for all participants younger than 19 years, starting insulin pump therapy before 31 December 2013. For each child, age, disease duration, date of insulin pump therapy initiation, insulin pump model, failure/malfunction/replacement yes/no and reason were considered for the year 2013. Results: Data were returned by 40 of 43 paediatric centres belonging to the Diabetes Study Group of the Italian Society of Paediatric Endocrinology and Diabetology. In total, 1574 of 11 311 (13.9%) children and adolescents with Type 1 diabetes were using an insulin pump: 29.2% Animas VIBE™, 9.4% Medtronic MiniMed 715/515™, 34.3% Medtronic MiniMed VEO™, 24.3% Accu-Check Spirit Combo™ and 2.8% other models. In 2013, 0.165 insulin pump replacements per patient-year (11.8% due to pump failure/malfunction and 4.7% due to accidental damage) were recorded. Animas VIBE™ (22.1%) and Medtronic MiniMed VEO™ (17.7%) were the most replaced. Conclusions: In a large cohort of Italian children and adolescents with Type 1 diabetes, insulin pump failure/malfunction and consequent replacement are aligned with rates previously reported and higher in more sophisticated pump models.

Insulin pump failures in Italian children with Type 1 diabetes : retrospective 1-year cohort study / I. Rabbone, N. Minuto, R. Bonfanti, M. Marigliano, F. Cerutti, V. Cherubini, G. D'Annunzio, A.P. Frongia, D. Iafusco, G. Ignaccolo, F. Lombardo, R. Schiaffini, S. Toni, S. Tumini, S. Zucchini, A. Pistorio, A.E. Scaramuzza, R. Lera, A. Secco, A. Bobbio, M. Bechaz, E. Piccinno, M.P. Natale, F. Ortolani, C. Zecchino, A. Lonero, G. Maltoni, B. Pasquino, F. Gallo, P. Frongia, C. Ripoli, D. Lo Presti, T. Timpanaro, F. Citriniti, T. Suprani, S. Carinci, P. Cipriano, N. Lazzaro, V. De Donno, F. Gallarotti, L. Lenzi, B. Piccini, L. Vittorio, C. Russo, R. Borea, F. Mammì, M. Bruzzese, C. Ventrici, G. Salzano, G. Frontino, C. Bonura, V. Favalli, A. Scaramuzza, G.V. Zuccotti, M. Ferrari, L. Iughetti, B. Predieri, A. Franzese, E. Mozzillo, P. Buono, S. Confetto, A. Zanfardino, F. Cadario, S. Savastio, C. Fiorito, P. Barbieri, G. Piredda, F. Cardella, R. Ropolo, G. Federico, B. Marchi, D. Benevento, C. Carducci, M.L. Mancabitti, M. Del Vecchio, R. Lapolla, A. Gaiero, G. Fichera, M.G. Ignaccolo, D. Tinti, V. Cauvin, R. Franceschi, M. Biagioni, A. Salvatoni, A. Scolari, C. Maffeis, A. Sabbion, C. Arnaldi, D. Tosini. - In: DIABETIC MEDICINE. - ISSN 0742-3071. - 34:5(2017), pp. 621-624. [10.1111/dme.13294]

Insulin pump failures in Italian children with Type 1 diabetes : retrospective 1-year cohort study

G.V. Zuccotti;
2017

Abstract

Aims: Insulin pump failure and/or malfunction requiring replacement have not been thoroughly investigated. This study evaluated pump replacement in children and adolescents with Type 1 diabetes using insulin pump therapy. Methods: Data were collected for all participants younger than 19 years, starting insulin pump therapy before 31 December 2013. For each child, age, disease duration, date of insulin pump therapy initiation, insulin pump model, failure/malfunction/replacement yes/no and reason were considered for the year 2013. Results: Data were returned by 40 of 43 paediatric centres belonging to the Diabetes Study Group of the Italian Society of Paediatric Endocrinology and Diabetology. In total, 1574 of 11 311 (13.9%) children and adolescents with Type 1 diabetes were using an insulin pump: 29.2% Animas VIBE™, 9.4% Medtronic MiniMed 715/515™, 34.3% Medtronic MiniMed VEO™, 24.3% Accu-Check Spirit Combo™ and 2.8% other models. In 2013, 0.165 insulin pump replacements per patient-year (11.8% due to pump failure/malfunction and 4.7% due to accidental damage) were recorded. Animas VIBE™ (22.1%) and Medtronic MiniMed VEO™ (17.7%) were the most replaced. Conclusions: In a large cohort of Italian children and adolescents with Type 1 diabetes, insulin pump failure/malfunction and consequent replacement are aligned with rates previously reported and higher in more sophisticated pump models.
Internal Medicine; Endocrinology, Diabetes and Metabolism; Endocrinology
Settore MED/38 - Pediatria Generale e Specialistica
2017
Article (author)
File in questo prodotto:
File Dimensione Formato  
Rabbone_et_al-2017-Diabetic_Medicine.pdf

accesso riservato

Tipologia: Publisher's version/PDF
Dimensione 59.85 kB
Formato Adobe PDF
59.85 kB Adobe PDF   Visualizza/Apri   Richiedi una copia
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/497920
Citazioni
  • ???jsp.display-item.citation.pmc??? 2
  • Scopus 12
  • ???jsp.display-item.citation.isi??? 10
social impact