RATIONALE: In the determination of immunosuppressive drugs cyclosporine A (CSA), tacrolimus (TARO), sirolimus (SIRO), and everolimus (EVE) in whole blood there is an open debate about which is the best assay between immunochemistry and liquid chromatography/tandem mass spectrometry (LC/MS/MS). This work is aimed to explore this topic, focusing on the use of updated assays and the analysis of a large number of samples. METHODS: A certified in vitro diagnostic kit coupled with a medical device LC/MS/MS was validated and applied to the analysis of 1192 blood samples of patients treated with immunosuppressive drugs. The results were compared with those obtained by immunoassays. RESULTS: The LC/MS/MS approach was found to provide linear, stable, precise, and accurate results, with lower limits of quantification of 12.5, 1.1, 1.2, and 1.2 μg/L for CSA, TACRO, SIRO, and EVE, respectively. With this method 80 samples were analysed and reported within a single work shift. A correlation was observed between the LC/MS/MS and immunoassay data, with Spearman correlation coefficients of 0.980 (n = 260) for CSA, 0.836 for TACRO (n = 562), 0.898 for SIRO (n = 113), and 0.904 for EVE (n = 257). Passing-Bablock regression showed the presence of constant and proportional biases for most of the drugs. A Blond-Altman graph showed differences between the assays, with immunoassays generally overestimating the drugs. CONCLUSIONS: The LC/MS/MS certified kit was validated for the detection of immunosuppressant drugs in whole blood and it provided a high-throughput method that is consistent with the requirements of clinical laboratories. The comparison of patient data between LC/MS/MS and up-dated immunoassays shows that a significant discrepancy still exists, especially for CSA and SIRO, confirming the greater specificity associated with use of the LC/MS/MS assay.

Immunosuppressive drugs in whole blood : validation of a commercially available LC-MS/MS kit and comparison with immunochemical assays / E. Polledri, R. Mercadante, C. Ferrari Fusarin, R. Maiavacca, S. Fustinoni. - In: RAPID COMMUNICATIONS IN MASS SPECTROMETRY. - ISSN 0951-4198. - 31:13(2017 Jul 15), pp. 1111-1120. [10.1002/rcm.7887]

Immunosuppressive drugs in whole blood : validation of a commercially available LC-MS/MS kit and comparison with immunochemical assays

R. Mercadante
Secondo
;
S. Fustinoni
Ultimo
2017

Abstract

RATIONALE: In the determination of immunosuppressive drugs cyclosporine A (CSA), tacrolimus (TARO), sirolimus (SIRO), and everolimus (EVE) in whole blood there is an open debate about which is the best assay between immunochemistry and liquid chromatography/tandem mass spectrometry (LC/MS/MS). This work is aimed to explore this topic, focusing on the use of updated assays and the analysis of a large number of samples. METHODS: A certified in vitro diagnostic kit coupled with a medical device LC/MS/MS was validated and applied to the analysis of 1192 blood samples of patients treated with immunosuppressive drugs. The results were compared with those obtained by immunoassays. RESULTS: The LC/MS/MS approach was found to provide linear, stable, precise, and accurate results, with lower limits of quantification of 12.5, 1.1, 1.2, and 1.2 μg/L for CSA, TACRO, SIRO, and EVE, respectively. With this method 80 samples were analysed and reported within a single work shift. A correlation was observed between the LC/MS/MS and immunoassay data, with Spearman correlation coefficients of 0.980 (n = 260) for CSA, 0.836 for TACRO (n = 562), 0.898 for SIRO (n = 113), and 0.904 for EVE (n = 257). Passing-Bablock regression showed the presence of constant and proportional biases for most of the drugs. A Blond-Altman graph showed differences between the assays, with immunoassays generally overestimating the drugs. CONCLUSIONS: The LC/MS/MS certified kit was validated for the detection of immunosuppressant drugs in whole blood and it provided a high-throughput method that is consistent with the requirements of clinical laboratories. The comparison of patient data between LC/MS/MS and up-dated immunoassays shows that a significant discrepancy still exists, especially for CSA and SIRO, confirming the greater specificity associated with use of the LC/MS/MS assay.
immunochemical assays; immunosuppressants in blood; liquid chromatography-tandem mass spectrometry
Settore MED/44 - Medicina del Lavoro
15-lug-2017
Article (author)
File in questo prodotto:
File Dimensione Formato  
immunosopressori_paper accettato.pdf

accesso riservato

Tipologia: Post-print, accepted manuscript ecc. (versione accettata dall'editore)
Dimensione 819.28 kB
Formato Adobe PDF
819.28 kB Adobe PDF   Visualizza/Apri   Richiedi una copia
Polledri_et_al-2017-Rapid_Communications_in_Mass_Spectrometry.pdf

accesso riservato

Tipologia: Publisher's version/PDF
Dimensione 600.68 kB
Formato Adobe PDF
600.68 kB Adobe PDF   Visualizza/Apri   Richiedi una copia
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/496913
Citazioni
  • ???jsp.display-item.citation.pmc??? 4
  • Scopus 17
  • ???jsp.display-item.citation.isi??? 17
social impact