Background The heel stick is the method of choice in most neonatal units for capillary blood sampling, and it represents the most common event among all painful procedures performed on newborns. The type and design of heel stick device and the clinical procedure to collect a blood sample may have an impact on newborn pain response as well. Objective To compare the pain response and efficiency of different automated devices for capillary blood collection in newborns. Design Randomized clinical trial. Setting Postnatal ward of a tertiary-care university hospital in Italy. Participants Newborn infants at gestational age ≥34 weeks undergoing the metabolic screening test after the 49th hour of life. Methods A total of 762 neonates were recruited and randomized into 6 groups (127 babies in each group) assigned to 6 different capillary blood collection devices (Ames Minilet™ Lancet; Cardinal Health Gentleheel®; Natus Medical NeatNick™; BD Quikheel™ Lancet; Vitrex Steriheel® Baby Lancet; Accriva Diagnostics Tenderfoot®). Main outcome measures The following data were collected and assessed for each of the 6 groups evaluated: a) number of heel sticks, b) pain score according to the Neonatal Infant Pain Scale (NIPS) and c) need to squeeze the heel. Results The Ames Minilet™ Lancet device was found to perform by far the worst compared to the five device underexamination: it required the highest number of sticks (mean = 3.91; 95% CI: 3.46–4.36), evoked the most intense pain (mean = 3.98; 95% CI: 3.77–4.20), and most frequently necessitated squeezing the heel (92.9%; 95% CI: 86.9–96.3). The five devices under examination appeared to be similar in terms of the number of sticks required, but differed slightly in NIPS score and in need to squeeze the heel. Conclusion The Accriva Diagnostics Tenderfoot® device demonstrated the greatest efficiency for blood sampling and evoked the least pain. With this device, the metabolic screening test could be performed with a single skin incision in the large majority of infants (98.4%), heel squeezing was limited to only 6.3% of infants, and the NIPS score turns out to be lower than other devices in our study (1.22; 95% CI 1.05–1.39).

The impact of automatic devices for capillary blood collection on efficiency and pain response in newborns: A randomized controlled trial / G. Sorrentino, M. Fumagalli, S. Milani, I. Cortinovis, A. Zorz, G. Cavallaro, F. Mosca, L. Plevani. - In: INTERNATIONAL JOURNAL OF NURSING STUDIES. - ISSN 0020-7489. - 72(2017), pp. 24-29. [10.1016/j.ijnurstu.2017.04.001]

The impact of automatic devices for capillary blood collection on efficiency and pain response in newborns: A randomized controlled trial

G. Sorrentino
;
M. Fumagalli
Secondo
;
S. Milani;I. Cortinovis;A. Zorz;G. Cavallaro;F. Mosca
Penultimo
;
L. Plevani
Ultimo
2017

Abstract

Background The heel stick is the method of choice in most neonatal units for capillary blood sampling, and it represents the most common event among all painful procedures performed on newborns. The type and design of heel stick device and the clinical procedure to collect a blood sample may have an impact on newborn pain response as well. Objective To compare the pain response and efficiency of different automated devices for capillary blood collection in newborns. Design Randomized clinical trial. Setting Postnatal ward of a tertiary-care university hospital in Italy. Participants Newborn infants at gestational age ≥34 weeks undergoing the metabolic screening test after the 49th hour of life. Methods A total of 762 neonates were recruited and randomized into 6 groups (127 babies in each group) assigned to 6 different capillary blood collection devices (Ames Minilet™ Lancet; Cardinal Health Gentleheel®; Natus Medical NeatNick™; BD Quikheel™ Lancet; Vitrex Steriheel® Baby Lancet; Accriva Diagnostics Tenderfoot®). Main outcome measures The following data were collected and assessed for each of the 6 groups evaluated: a) number of heel sticks, b) pain score according to the Neonatal Infant Pain Scale (NIPS) and c) need to squeeze the heel. Results The Ames Minilet™ Lancet device was found to perform by far the worst compared to the five device underexamination: it required the highest number of sticks (mean = 3.91; 95% CI: 3.46–4.36), evoked the most intense pain (mean = 3.98; 95% CI: 3.77–4.20), and most frequently necessitated squeezing the heel (92.9%; 95% CI: 86.9–96.3). The five devices under examination appeared to be similar in terms of the number of sticks required, but differed slightly in NIPS score and in need to squeeze the heel. Conclusion The Accriva Diagnostics Tenderfoot® device demonstrated the greatest efficiency for blood sampling and evoked the least pain. With this device, the metabolic screening test could be performed with a single skin incision in the large majority of infants (98.4%), heel squeezing was limited to only 6.3% of infants, and the NIPS score turns out to be lower than other devices in our study (1.22; 95% CI 1.05–1.39).
Blood specimen collection; Capillary; Heel; Newborn; Pain; Puncture; Nursing (all)2901 Nursing (miscellaneous)
Settore MED/38 - Pediatria Generale e Specialistica
2017
www.elsevier.com/locate/ijnurstu
Article (author)
File in questo prodotto:
File Dimensione Formato  
The impact of automatic devices for capillary blood collection on efficiency.pdf

accesso riservato

Tipologia: Post-print, accepted manuscript ecc. (versione accettata dall'editore)
Dimensione 461.35 kB
Formato Adobe PDF
461.35 kB Adobe PDF   Visualizza/Apri   Richiedi una copia
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/495199
Citazioni
  • ???jsp.display-item.citation.pmc??? 2
  • Scopus 10
  • ???jsp.display-item.citation.isi??? 8
social impact