The reference measurement procedure for ALP published in 2011 by the IFCC allowed to define the metrological traceability chain for the standardization of ALP measurement. This paper reports the results of an EQA experiment conducted to evaluate the level of ALP standardization among different Italian laboratories enrolled for a scientific project with the final aim to derive ALP traceable reference intervals for pediatric population. Three frozen serum pools with a target value assigned by the IFCC reference procedure were distributed to 13 centers and analyzed in triplicates for 3 different days. Only 3 laboratories averagely fulfilled the desirable goal of bias (≤±5.5%) at all 3 concentrations (59.9 U/L, 186.9 U/L and 401.5 U/L), but only one provided data with a dispersion always within the uncertainty of the target result. The different ability to meet the goal clearly depended on the analytical system used. Focusing on the two most used analytical platforms, the Cobas systems (Roche Diagnostics) underestimated the ALP values, while the AU systems (Beckman Coulter) overestimated them. The regression parameters between the average values obtained by laboratories and the target values indicate that it would be possible to correct the results of all analytical systems and make them unbiased by a simple recalibration approach. The analysis of the commercial calibrator package inserts of the IVD companies involved in this study showed that, with the exception of Roche still aligned to the old Tietz method published in 1983, all companies offer at least two options, sometimes (e.g., Beckman AU) both not in line with the recommended standardization approach.

Valutazione dell’esattezza della misura della fosfatasi alcalina sierica in un gruppo di laboratori italiani = Evaluation of the trueness of serum alkaline phosphatase (ALP) measurement in a group of Italian laboratories / F. Braga, E. Frusciante, I. Infusino, E. Aloisio, E. Guerra, F. Ceriotti, M. Panteghini. - In: BIOCHIMICA CLINICA. - ISSN 0393-0564. - 41:1(2017), pp. 64-71. [10.19186/BC-2017.005]

Valutazione dell’esattezza della misura della fosfatasi alcalina sierica in un gruppo di laboratori italiani = Evaluation of the trueness of serum alkaline phosphatase (ALP) measurement in a group of Italian laboratories

F. Braga;M. Panteghini
2017

Abstract

The reference measurement procedure for ALP published in 2011 by the IFCC allowed to define the metrological traceability chain for the standardization of ALP measurement. This paper reports the results of an EQA experiment conducted to evaluate the level of ALP standardization among different Italian laboratories enrolled for a scientific project with the final aim to derive ALP traceable reference intervals for pediatric population. Three frozen serum pools with a target value assigned by the IFCC reference procedure were distributed to 13 centers and analyzed in triplicates for 3 different days. Only 3 laboratories averagely fulfilled the desirable goal of bias (≤±5.5%) at all 3 concentrations (59.9 U/L, 186.9 U/L and 401.5 U/L), but only one provided data with a dispersion always within the uncertainty of the target result. The different ability to meet the goal clearly depended on the analytical system used. Focusing on the two most used analytical platforms, the Cobas systems (Roche Diagnostics) underestimated the ALP values, while the AU systems (Beckman Coulter) overestimated them. The regression parameters between the average values obtained by laboratories and the target values indicate that it would be possible to correct the results of all analytical systems and make them unbiased by a simple recalibration approach. The analysis of the commercial calibrator package inserts of the IVD companies involved in this study showed that, with the exception of Roche still aligned to the old Tietz method published in 1983, all companies offer at least two options, sometimes (e.g., Beckman AU) both not in line with the recommended standardization approach.
Settore BIO/12 - Biochimica Clinica e Biologia Molecolare Clinica
2017
Centro per la riferibilità Metrologica in Medicina di Laboratorio (CIRME)
Article (author)
File in questo prodotto:
File Dimensione Formato  
1304-064-071-braga.pdf

accesso riservato

Tipologia: Publisher's version/PDF
Dimensione 193.35 kB
Formato Adobe PDF
193.35 kB Adobe PDF   Visualizza/Apri   Richiedi una copia
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/487253
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 1
  • ???jsp.display-item.citation.isi??? ND
social impact