A thorough evaluation of the dose inside a specially designed and built facility for extra-corporeal treatment of liver cancer by boron neutron capture therapy (BNCT) at the High Flux Reactor (HFR) Petten (The Netherlands) is the necessary step before animal studies can start. The absorbed doses are measured by means of gel dosemeters, which help to validate the Monte Carlo simulations of the spheroidal liver holder that will contain the human liver for irradiation with an epithermal neutron beam. These dosemeters allow imaging of the dose due to gammas and to the charged particles produced by the 10B reaction. The thermal neutron flux is extrapolated from the boron dose images and compared to that obtained by the calculations. As an additional reference, Au, Cu and Mn foil measurements are performed. All results appear consistent with the calculations and confirm that the BNCT liver facility is able to provide an almost homogeneous thermal neutron distribution in the liver, which is a requirement for a successful treatment of liver metastases.

Gel dosimetry in the BNCT facility for extra-corporeal treatment of liver cancer at the HFR Petten / G. Gambarini, G.G. Daquino, R.L. Moss, M. Carrara, V.A. Nievaart, E. Vanossi. - In: RADIATION PROTECTION DOSIMETRY. - ISSN 0144-8420. - 126:1-4(2007), pp. 604-609.

Gel dosimetry in the BNCT facility for extra-corporeal treatment of liver cancer at the HFR Petten

G. Gambarini
Primo
;
M. Carrara;
2007

Abstract

A thorough evaluation of the dose inside a specially designed and built facility for extra-corporeal treatment of liver cancer by boron neutron capture therapy (BNCT) at the High Flux Reactor (HFR) Petten (The Netherlands) is the necessary step before animal studies can start. The absorbed doses are measured by means of gel dosemeters, which help to validate the Monte Carlo simulations of the spheroidal liver holder that will contain the human liver for irradiation with an epithermal neutron beam. These dosemeters allow imaging of the dose due to gammas and to the charged particles produced by the 10B reaction. The thermal neutron flux is extrapolated from the boron dose images and compared to that obtained by the calculations. As an additional reference, Au, Cu and Mn foil measurements are performed. All results appear consistent with the calculations and confirm that the BNCT liver facility is able to provide an almost homogeneous thermal neutron distribution in the liver, which is a requirement for a successful treatment of liver metastases.
English
BNCT ; dose imaging ; neutron dosimetry ; gel dosimetry
Settore FIS/01 - Fisica Sperimentale
Articolo
Esperti anonimi
2007
Oxford University Press
126
1-4
604
609
Periodico con rilevanza internazionale
info:eu-repo/semantics/article
Gel dosimetry in the BNCT facility for extra-corporeal treatment of liver cancer at the HFR Petten / G. Gambarini, G.G. Daquino, R.L. Moss, M. Carrara, V.A. Nievaart, E. Vanossi. - In: RADIATION PROTECTION DOSIMETRY. - ISSN 0144-8420. - 126:1-4(2007), pp. 604-609.
none
Prodotti della ricerca::01 - Articolo su periodico
6
262
Article (author)
si
G. Gambarini, G.G. Daquino, R.L. Moss, M. Carrara, V.A. Nievaart, E. Vanossi
File in questo prodotto:
Non ci sono file associati a questo prodotto.
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/38367
Citazioni
  • ???jsp.display-item.citation.pmc??? 0
  • Scopus 6
  • ???jsp.display-item.citation.isi??? 3
social impact