BACKGROUND: The purpose of this study was to evaluate if there is a difference in the rates of infection between Ilizarov wires and half-pins and between half-pins with threads outside and inside the skin in circular fixators modified by Catagni and Cattaneo. METHODS: Between May and December 2004, 218 patients with circular Ilizarov fixators for various orthopaedic indications who visited our Ilizarov clinic were subjected to a one-time evaluation of half-pins and wires. RESULTS: A total of 1,093 half-pins were checked and 34 (3.11%) of these were infected (50% with threads outside and 50% with threads inside the skin). Among a total number of 951 wires (1,092 pin tracts) checked, 45 (4.73%) were infected. CONCLUSIONS: We conclude that the circular Ilizarov fixation with conical half-pins and wires has similar rate of infection compared with conventional Ilizarov circular fixator. Moreover, there is no difference in infection rates between pins with threads inside the skin as compared to those with threads outside the skin.

External circular fixation: a comparison of infection rates between wires and conical half-pins with threads outside or inside the skin / M.A. Catagni, G. Ottaviani, A. Combi, A. Elhence. - In: THE JOURNAL OF TRAUMA, INJURY, INFECTION, AND CRITICAL CARE. - ISSN 0022-5282. - 61:5(2006 Nov), pp. 1186-1191.

External circular fixation: a comparison of infection rates between wires and conical half-pins with threads outside or inside the skin

G. Ottaviani
Secondo
;
A. Combi
Penultimo
;
2006

Abstract

BACKGROUND: The purpose of this study was to evaluate if there is a difference in the rates of infection between Ilizarov wires and half-pins and between half-pins with threads outside and inside the skin in circular fixators modified by Catagni and Cattaneo. METHODS: Between May and December 2004, 218 patients with circular Ilizarov fixators for various orthopaedic indications who visited our Ilizarov clinic were subjected to a one-time evaluation of half-pins and wires. RESULTS: A total of 1,093 half-pins were checked and 34 (3.11%) of these were infected (50% with threads outside and 50% with threads inside the skin). Among a total number of 951 wires (1,092 pin tracts) checked, 45 (4.73%) were infected. CONCLUSIONS: We conclude that the circular Ilizarov fixation with conical half-pins and wires has similar rate of infection compared with conventional Ilizarov circular fixator. Moreover, there is no difference in infection rates between pins with threads inside the skin as compared to those with threads outside the skin.
Catagni-Cattaneo circular fixator; Conical half-pins; Ilizarov method; Infection rates; Kirschner wires; Pin-tract infection
Settore MED/08 - Anatomia Patologica
Settore MED/33 - Malattie Apparato Locomotore
nov-2006
Centro di Ricerca "Lino Rossi" per lo studio e la prevenzione della morte perinatale del feto a termine e neonatale e della sindrome della morte improvvisa del lattante SIDS
Article (author)
File in questo prodotto:
File Dimensione Formato  
128_JT_06.pdf

accesso riservato

Tipologia: Publisher's version/PDF
Dimensione 1.8 MB
Formato Adobe PDF
1.8 MB Adobe PDF   Visualizza/Apri   Richiedi una copia
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/29943
Citazioni
  • ???jsp.display-item.citation.pmc??? 2
  • Scopus 17
  • ???jsp.display-item.citation.isi??? 15
social impact