The huge amount of optical photons emitted in burst by Cerium doped Lanthanum bromide scintillators poses a severe constraint on the specifications of photomultipliers, namely they must operate at very large instantaneous current values with minimum saturation of output pulse signals. Acquisition in shape and processing of photomultipliers output pulses by means of high-speed free-running analog to digital converters and processing devices allows specific correction for pulse saturation and distortion effects, due to high-gain tubes operated at relatively high power supply voltage. We analyze the effects of output signal saturation in photomultiplier tubes as a function of the high voltage level and the amount of energy released in the scintillator crystal, proposing a simple model to characterize this complex phenomenon. According to the proposed model, a preliminary first-order correction is calculated and applied to the energy estimations obtained by a traditional analog spectroscopy system, for energies up to 9 MeV, reducing the maximum estimation error by almost an order of magnitude.

Analysis and first order correction of signal saturation effects in photomultiplier tubes for improved estimation of interacting radiation energy in Lanthanum bromide scintillators / N. Blasi, S. Brambilla, C. Boiano, F. Camera, A. Camplani, F.C.L. Crespi, A. Giaz, B. Million, R. Nicolini, L. Pellegri, S. Riboldi, O. Wieland - In: Nuclear Science Symposium Conference Record (NSS/MIC), 2010 IEEE : Oct. 30 2010-Nov. 6 2010[s.l] : IEEE, 2010. - ISBN 9781424491063. - pp. 1809-1812 (( convegno IEEE Nuclear Science Symposium, Medical Imaging Conference tenutosi a Knoxville nel 2010 [10.1109/NSSMIC.2010.5874087].

Analysis and first order correction of signal saturation effects in photomultiplier tubes for improved estimation of interacting radiation energy in Lanthanum bromide scintillators

F. Camera;A. Camplani;F.C.L. Crespi;A. Giaz;L. Pellegri;S. Riboldi;
2010

Abstract

The huge amount of optical photons emitted in burst by Cerium doped Lanthanum bromide scintillators poses a severe constraint on the specifications of photomultipliers, namely they must operate at very large instantaneous current values with minimum saturation of output pulse signals. Acquisition in shape and processing of photomultipliers output pulses by means of high-speed free-running analog to digital converters and processing devices allows specific correction for pulse saturation and distortion effects, due to high-gain tubes operated at relatively high power supply voltage. We analyze the effects of output signal saturation in photomultiplier tubes as a function of the high voltage level and the amount of energy released in the scintillator crystal, proposing a simple model to characterize this complex phenomenon. According to the proposed model, a preliminary first-order correction is calculated and applied to the energy estimations obtained by a traditional analog spectroscopy system, for energies up to 9 MeV, reducing the maximum estimation error by almost an order of magnitude.
No
English
analogue-digital conversion;germanium radiation detectors; nuclear electronics; photomultipliers; solid scintillation detectors; analog spectroscopy system; free-running analog-to-digital converters; high-gain tubes; instantaneous current values; interacting radiation energy; lanthanum bromide scintillators; optical photons; output pulse signals; photomultiplier tubes; photomultipliers output pulses; power supply voltage; scintillator crystal; signal saturation effects
Settore FIS/04 - Fisica Nucleare e Subnucleare
Intervento a convegno
Sì, ma tipo non specificato
Ricerca di base
Pubblicazione scientifica
Nuclear Science Symposium Conference Record (NSS/MIC), 2010 IEEE : Oct. 30 2010-Nov. 6 2010
IEEE
2010
1809
1812
4
9781424491063
Volume a diffusione internazionale
IEEE Nuclear Science Symposium, Medical Imaging Conference
Knoxville
2010
Nucl. Plasma Sci. Soc. Inst. Electr. Electron. Eng. (NPSS)
Convegno internazionale
Intervento inviato
scopus
crossref
Aderisco
N. Blasi, S. Brambilla, C. Boiano, F. Camera, A. Camplani, F.C.L. Crespi, A. Giaz, B. Million, R. Nicolini, L. Pellegri, S. Riboldi, O. Wieland...espandi
Book Part (author)
reserved
273
Analysis and first order correction of signal saturation effects in photomultiplier tubes for improved estimation of interacting radiation energy in Lanthanum bromide scintillators / N. Blasi, S. Brambilla, C. Boiano, F. Camera, A. Camplani, F.C.L. Crespi, A. Giaz, B. Million, R. Nicolini, L. Pellegri, S. Riboldi, O. Wieland - In: Nuclear Science Symposium Conference Record (NSS/MIC), 2010 IEEE : Oct. 30 2010-Nov. 6 2010[s.l] : IEEE, 2010. - ISBN 9781424491063. - pp. 1809-1812 (( convegno IEEE Nuclear Science Symposium, Medical Imaging Conference tenutosi a Knoxville nel 2010 [10.1109/NSSMIC.2010.5874087].
info:eu-repo/semantics/bookPart
12
Prodotti della ricerca::03 - Contributo in volume
File in questo prodotto:
File Dimensione Formato  
05874087.pdf

accesso riservato

Tipologia: Publisher's version/PDF
Dimensione 1.35 MB
Formato Adobe PDF
1.35 MB Adobe PDF   Visualizza/Apri   Richiedi una copia
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/248376
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 1
  • ???jsp.display-item.citation.isi??? 0
  • OpenAlex ND
social impact