Purpose: To evaluate the marginal adaptation at implant-abutment connection of an implant featuring a conical (45° taper) internal hexagonal abutment with a connection depth of 2.5. mm, comparing the performance of two identical abutments of different material (titanium grade-4 and Co-Cr-alloy). Methods: Twenty implants (3.75. mm × 15. mm) were connected to non-matching abutments (5.5. mm × 10. mm) of two different materials (titanium grade-4: n = 10; Co-Cr-alloy: n = 10). The specimens were separately embedded in epoxylite resin, inside copper cylinders, and submerged without covering the most coronal portion (5. mm) of the fixture. Five specimens per group were stressed simulating a surgical 100. Ncm insertion torque, while the others had no torque simulation. All specimens were subjected to a non-axial static load (100. N) in a universal testing machine, under an angle of 30° with respect to the implant axis. Once 100. N load was reached, low shrinkage self-curing resin was injected inside the cylinders, and load was maintained until complete resin polymerization. Specimens were cut and analyzed with optical and scanning-electron-microscope (SEM) to evaluate the marginal adaptation at the implant-abutment connection. Statistical analysis was performed using one-way ANOVA (p = 0.02). Results: None of the 20 samples failed. The implant-abutment connection was able to guarantee a good optical seal; SEM analysis confirmed the absence of microgaps. Conclusions: Within the limits of this study (small sample size, limited time) the marginal adaptation of the implant-abutment connection was not affected by the abutment material nor by the application of surgical insertion torque.

Effects of different abutment material and surgical insertion torque on the marginal adaptation of an internal conical interface: An in vitro study / D. Farronato, S. Pieroni, F.G. Mangano, F. Briguglio, D. Re. - In: JOURNAL OF PROSTHODONTIC RESEARCH. - ISSN 1883-1958. - 58:4(2014 Sep), pp. 230-236. [Epub ahead of print] [10.1016/j.jpor.2014.05.004.]

Effects of different abutment material and surgical insertion torque on the marginal adaptation of an internal conical interface: An in vitro study

D. Re
Primo
2014

Abstract

Purpose: To evaluate the marginal adaptation at implant-abutment connection of an implant featuring a conical (45° taper) internal hexagonal abutment with a connection depth of 2.5. mm, comparing the performance of two identical abutments of different material (titanium grade-4 and Co-Cr-alloy). Methods: Twenty implants (3.75. mm × 15. mm) were connected to non-matching abutments (5.5. mm × 10. mm) of two different materials (titanium grade-4: n = 10; Co-Cr-alloy: n = 10). The specimens were separately embedded in epoxylite resin, inside copper cylinders, and submerged without covering the most coronal portion (5. mm) of the fixture. Five specimens per group were stressed simulating a surgical 100. Ncm insertion torque, while the others had no torque simulation. All specimens were subjected to a non-axial static load (100. N) in a universal testing machine, under an angle of 30° with respect to the implant axis. Once 100. N load was reached, low shrinkage self-curing resin was injected inside the cylinders, and load was maintained until complete resin polymerization. Specimens were cut and analyzed with optical and scanning-electron-microscope (SEM) to evaluate the marginal adaptation at the implant-abutment connection. Statistical analysis was performed using one-way ANOVA (p = 0.02). Results: None of the 20 samples failed. The implant-abutment connection was able to guarantee a good optical seal; SEM analysis confirmed the absence of microgaps. Conclusions: Within the limits of this study (small sample size, limited time) the marginal adaptation of the implant-abutment connection was not affected by the abutment material nor by the application of surgical insertion torque.
No
English
Dental implants; Implant-abutment connection; Marginal adaptation; Microgap
Settore MED/28 - Malattie Odontostomatologiche
Articolo
Esperti anonimi
Pubblicazione scientifica
set-2014
Elsevier
58
4
230
236
7
Epub ahead of print
Periodico con rilevanza internazionale
info:eu-repo/semantics/article
Effects of different abutment material and surgical insertion torque on the marginal adaptation of an internal conical interface: An in vitro study / D. Farronato, S. Pieroni, F.G. Mangano, F. Briguglio, D. Re. - In: JOURNAL OF PROSTHODONTIC RESEARCH. - ISSN 1883-1958. - 58:4(2014 Sep), pp. 230-236. [Epub ahead of print] [10.1016/j.jpor.2014.05.004.]
reserved
Prodotti della ricerca::01 - Articolo su periodico
5
262
Article (author)
no
D. Farronato, S. Pieroni, F.G. Mangano, F. Briguglio, D. Re
File in questo prodotto:
File Dimensione Formato  
1 Effect of JPRSR 2014.pdf

accesso riservato

Tipologia: Publisher's version/PDF
Dimensione 1.33 MB
Formato Adobe PDF
1.33 MB Adobe PDF   Visualizza/Apri   Richiedi una copia
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/238969
Citazioni
  • ???jsp.display-item.citation.pmc??? 2
  • Scopus 4
  • ???jsp.display-item.citation.isi??? 4
social impact