Appropriate limb length (LL) and femoral offset (OS) after total hip arthroplasty (THA) are crucial for a successful and lasting outcome. However, intraoperative assessment is difficult. Computed navigation is effective but the use of radiation and of invasive bone references is cause of concern. Imageless, pinless navigation systems have been shown to be accurate in experimental studies. However, almost no clinical validation has been performed. We used a minimally-invasive computer-navigated system (MICNS) in conjunction with an imageless measurement technique for implanting 60 consecutive THAs. Head/neck trial components of different size were applied, and the differences in LL and femoral OS measurements were recorded and compared to the implant manufacturer's specifications, corrected for the anatomical variations of each patient. The imageless MICNS revealed a valid and accurate intra-operative tool for measuring LL and femoral OS in vivo.

Intraoperative validation of navigated limb measurements in THA using a pinless femoral array / M. Ulivi, L. Orlandini, W. Pascale, O. Consonni, V. Sansone. - In: THE JOURNAL OF ARTHROPLASTY. - ISSN 0883-5403. - 29:5(2014 May), pp. 1026-1029.

Intraoperative validation of navigated limb measurements in THA using a pinless femoral array

O. Consonni;V. Sansone
2014

Abstract

Appropriate limb length (LL) and femoral offset (OS) after total hip arthroplasty (THA) are crucial for a successful and lasting outcome. However, intraoperative assessment is difficult. Computed navigation is effective but the use of radiation and of invasive bone references is cause of concern. Imageless, pinless navigation systems have been shown to be accurate in experimental studies. However, almost no clinical validation has been performed. We used a minimally-invasive computer-navigated system (MICNS) in conjunction with an imageless measurement technique for implanting 60 consecutive THAs. Head/neck trial components of different size were applied, and the differences in LL and femoral OS measurements were recorded and compared to the implant manufacturer's specifications, corrected for the anatomical variations of each patient. The imageless MICNS revealed a valid and accurate intra-operative tool for measuring LL and femoral OS in vivo.
accuracy; computer navigation; hip arthroplasty; pinless femoral array
Settore MED/33 - Malattie Apparato Locomotore
mag-2014
Article (author)
File in questo prodotto:
File Dimensione Formato  
2014 Intraoperative Validation of Navigated Limb Measurements in THA Using a Pinless Femoral Array.pdf

accesso riservato

Tipologia: Publisher's version/PDF
Dimensione 582.45 kB
Formato Adobe PDF
582.45 kB Adobe PDF   Visualizza/Apri   Richiedi una copia
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/237166
Citazioni
  • ???jsp.display-item.citation.pmc??? 1
  • Scopus 7
  • ???jsp.display-item.citation.isi??? 3
social impact