Radiation detectors on a semi-insulating (SI) 4H silicon carbide (SiC) wafer have been manufactured and characterized with X and γ photons in the range 8-59 keV. The detectors were 400-μm-diameter circular Ni-SiC junctions on an SI 4H-SiC wafer thinned to 70 μm. Dark current densities of 3.5 nA/cm2 at +20°C and 0.3 μA / cm2 at +104°C with an internal electric field of 7 kV/cm have been measured. X-γ ray spectra from 241Am have been acquired at room temperature with pulser line width of 756 eV FWHM. The charge collection efficiency (CCE) has been measured under different experimental conditions with a maximum CCE = 75% at room temperature. Polarization effects have been observed, and the dependence of CCE on time and temperature has been measured and analyzed. The charge trapping has been described by the Hecht model with a maximum total mean drift length of 107 μm at room temperature.

X- γ ray spectroscopy with semi-insulating 4H-silicon carbide / G. Bertuccio, D. Puglisi, A. Pullia, C. Lanzieri. - In: IEEE TRANSACTIONS ON NUCLEAR SCIENCE. - ISSN 0018-9499. - 60:2(2013 Apr), pp. 6496305.1436-6496305.1441.

X- γ ray spectroscopy with semi-insulating 4H-silicon carbide

A. Pullia;
2013

Abstract

Radiation detectors on a semi-insulating (SI) 4H silicon carbide (SiC) wafer have been manufactured and characterized with X and γ photons in the range 8-59 keV. The detectors were 400-μm-diameter circular Ni-SiC junctions on an SI 4H-SiC wafer thinned to 70 μm. Dark current densities of 3.5 nA/cm2 at +20°C and 0.3 μA / cm2 at +104°C with an internal electric field of 7 kV/cm have been measured. X-γ ray spectra from 241Am have been acquired at room temperature with pulser line width of 756 eV FWHM. The charge collection efficiency (CCE) has been measured under different experimental conditions with a maximum CCE = 75% at room temperature. Polarization effects have been observed, and the dependence of CCE on time and temperature has been measured and analyzed. The charge trapping has been described by the Hecht model with a maximum total mean drift length of 107 μm at room temperature.
Charge collection efficiency (CCE); polarization effect; radiation detectors; semi-insulating (SI) crystals; semiconductor detectors; silicon carbide (SiC); X-ray spectroscopy
Settore ING-INF/01 - Elettronica
Settore FIS/01 - Fisica Sperimentale
apr-2013
Article (author)
File in questo prodotto:
File Dimensione Formato  
X-gamma Ray Spectroscopy With Semi-Insulating 4H-Silicon Carbide.pdf

accesso riservato

Tipologia: Publisher's version/PDF
Dimensione 1 MB
Formato Adobe PDF
1 MB Adobe PDF   Visualizza/Apri   Richiedi una copia
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/232872
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 27
  • ???jsp.display-item.citation.isi??? 24