The properties of large volume cylindrical 3.5 '' x 8 '' (89 mm x 203 mm) LaBr3:Ce scintillation detectors coupled to the Hamamatsu R10233-100SEL photo-multiplier tube were investigated. These crystals are among the largest ones ever produced and still need to be fully characterized to determine how these detectors can be utilized and in which applications. We tested the detectors using monochromatic gamma-ray sources and in-beam reactions producing gamma rays up to 22.6 MeV; we acquired PMT signal pulses and calculated detector energy resolution and response linearity as a function of gamma-ray energy. Two different voltage dividers were coupled to the Hamamatsu R10233-100SEL PMT: the Hamamatsu E1198-26, based on straightforward resistive network design, and the "LABRVD", specifically designed for our large volume LaBr3:Ce scintillation detectors, which also includes active semiconductor devices. Because of the extremely high light yield of LaBr3:Ce crystals we observed that depending on the choice of PMT, voltage divider and applied voltage, some significant deviation from the ideally proportional response of the detector and some pulse shape deformation appear. In addition, crystal non-homogeneities and PMT gain drifts affect the (measured) energy resolution especially in case of high-energy gamma rays. We also measured the time resolution of detectors with different sizes (from 1 '' x 1 '' up to 3.5 '' x 8 ''), correlating the results with both the intrinsic properties of PMTs and (WANT simulations of the scintillation light collection process. The detector absolute full energy efficiency was measured and simulated up to gamma-rays of 30 MeV.

Characterization of large volume 3.5 '' x 8 '' LaBr3:Ce detectors / A. Giaz, L. Pellegri, S. Riboldi, F. Camera, N. Blasi, C. Boiano, A. Bracco, S. Brambilla, S. Ceruti, S. Coelli, F.C.L. Crespi, M. Csatlos, S. Frega, J. Gulyas, A. Krasznahorkay, S. Lodetti, B. Million, A. Owens, F. Quarati, L. Stuhl, O. Wieland. - In: NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH. SECTION A, ACCELERATORS, SPECTROMETERS, DETECTORS AND ASSOCIATED EQUIPMENT. - ISSN 0168-9002. - 729:(2013), pp. 910-921. [10.1016/j.nima.2013.07.084]

Characterization of large volume 3.5 '' x 8 '' LaBr3:Ce detectors

A. Giaz;L. Pellegri;S. Riboldi;F. Camera;A. Bracco;S. Ceruti;F.C.L. Crespi;
2013

Abstract

The properties of large volume cylindrical 3.5 '' x 8 '' (89 mm x 203 mm) LaBr3:Ce scintillation detectors coupled to the Hamamatsu R10233-100SEL photo-multiplier tube were investigated. These crystals are among the largest ones ever produced and still need to be fully characterized to determine how these detectors can be utilized and in which applications. We tested the detectors using monochromatic gamma-ray sources and in-beam reactions producing gamma rays up to 22.6 MeV; we acquired PMT signal pulses and calculated detector energy resolution and response linearity as a function of gamma-ray energy. Two different voltage dividers were coupled to the Hamamatsu R10233-100SEL PMT: the Hamamatsu E1198-26, based on straightforward resistive network design, and the "LABRVD", specifically designed for our large volume LaBr3:Ce scintillation detectors, which also includes active semiconductor devices. Because of the extremely high light yield of LaBr3:Ce crystals we observed that depending on the choice of PMT, voltage divider and applied voltage, some significant deviation from the ideally proportional response of the detector and some pulse shape deformation appear. In addition, crystal non-homogeneities and PMT gain drifts affect the (measured) energy resolution especially in case of high-energy gamma rays. We also measured the time resolution of detectors with different sizes (from 1 '' x 1 '' up to 3.5 '' x 8 ''), correlating the results with both the intrinsic properties of PMTs and (WANT simulations of the scintillation light collection process. The detector absolute full energy efficiency was measured and simulated up to gamma-rays of 30 MeV.
Energy resolution; Gamma rays; LaBr3:Ce; Large volume scintillator detectors; Time resoluttion
Settore FIS/04 - Fisica Nucleare e Subnucleare
2013
Article (author)
File in questo prodotto:
File Dimensione Formato  
1-s2.0-S016890021301111X-main.pdf

accesso solo dalla rete interna

Tipologia: Publisher's version/PDF
Dimensione 1.17 MB
Formato Adobe PDF
1.17 MB Adobe PDF   Visualizza/Apri   Richiedi una copia
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/229821
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 98
  • ???jsp.display-item.citation.isi??? 94
social impact