Purpose: To evaluate the ability of frequency-doubling technology (FDT) perimetry in detecting glaucoma with N-30 and C-20 screening programs. Methods: Eighty eyes of 80 patients were enrolled (40 glaucomatous, 40 controls). Humphrey achromatic perimetry (standard automated perimetry, SAP) was considered as the "gold standard" for diagnosis. To assess whether N-30 screening program could detect more initial glaucomatous defects than C-20, glaucomatous patients included 20 cases with nasal step at SAP (pre-selected by medical chart inspection). Patients underwent two SAP examinations to confirm diagnosis; then two N-30 and two C-20 screening tests with frequency-doubling technology were performed in a randomized sequence. Finally, a frequency-doubling technology N-30 full-threshold examination was performed. Several criteria to define abnormality at frequency-doubling technology screening programs were evaluated. Results: For both C-20 and N-30 screening programs, the best parameter to detect glaucoma was the presence of at least 1 point with P < 5% (sensitivity = 87.5% for both tests and specificity of 90% and 95% for C-20 and N-30, respectively). Both screening procedures obtained a lower sensitivity (75%) in patients with a nasal step, whereas frequency-doubling technology full-threshold program was able to detect the initial defects in all cases. Conclusions: N-30 and C-20 screening procedures obtained similar results in well-defined glaucoma patients in terms of sensitivity and specificity. In the presence of a standard automated perimetry nasal step, diagnostic ability with both frequency-doubling technology screening strategies decreased and one quarter of nasal steps went undetected.

Detecting glaucoma with frequency-doubling technology perimetry : a comparison between N-30 and C-20 screening programs / P. Fogagnolo, F. Mazzolani, L. Rossetti, N. Orzalesi. - In: JOURNAL OF GLAUCOMA. - ISSN 1057-0829. - 14:6(2005 Dec), pp. 485-491.

Detecting glaucoma with frequency-doubling technology perimetry : a comparison between N-30 and C-20 screening programs

P. Fogagnolo
Primo
;
F. Mazzolani
Secondo
;
L. Rossetti
Penultimo
;
N. Orzalesi
Ultimo
2005

Abstract

Purpose: To evaluate the ability of frequency-doubling technology (FDT) perimetry in detecting glaucoma with N-30 and C-20 screening programs. Methods: Eighty eyes of 80 patients were enrolled (40 glaucomatous, 40 controls). Humphrey achromatic perimetry (standard automated perimetry, SAP) was considered as the "gold standard" for diagnosis. To assess whether N-30 screening program could detect more initial glaucomatous defects than C-20, glaucomatous patients included 20 cases with nasal step at SAP (pre-selected by medical chart inspection). Patients underwent two SAP examinations to confirm diagnosis; then two N-30 and two C-20 screening tests with frequency-doubling technology were performed in a randomized sequence. Finally, a frequency-doubling technology N-30 full-threshold examination was performed. Several criteria to define abnormality at frequency-doubling technology screening programs were evaluated. Results: For both C-20 and N-30 screening programs, the best parameter to detect glaucoma was the presence of at least 1 point with P < 5% (sensitivity = 87.5% for both tests and specificity of 90% and 95% for C-20 and N-30, respectively). Both screening procedures obtained a lower sensitivity (75%) in patients with a nasal step, whereas frequency-doubling technology full-threshold program was able to detect the initial defects in all cases. Conclusions: N-30 and C-20 screening procedures obtained similar results in well-defined glaucoma patients in terms of sensitivity and specificity. In the presence of a standard automated perimetry nasal step, diagnostic ability with both frequency-doubling technology screening strategies decreased and one quarter of nasal steps went undetected.
Frequency-doubling technology perimetry; Glaucoma; Nasal step; Screening; Standard automated perimetry
Settore MED/30 - Malattie Apparato Visivo
dic-2005
Article (author)
File in questo prodotto:
File Dimensione Formato  
00061198-200512000-00013.pdf

accesso riservato

Tipologia: Publisher's version/PDF
Dimensione 96.95 kB
Formato Adobe PDF
96.95 kB Adobe PDF   Visualizza/Apri   Richiedi una copia
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/22692
Citazioni
  • ???jsp.display-item.citation.pmc??? 3
  • Scopus 18
  • ???jsp.display-item.citation.isi??? 16
social impact