Purpose:To assess repeatability (intra-observer variability) and reproducibility (inter-observer variability) of intraocular pressure (IOP) measurements with Applanation Resonance Tonometry (ART) and to evaluate possible influential factors. Methods:The study included 178 patients (101 glaucoma, 63 healthy and 14 with OHT; one eye per subject) from two Italian centers (Eye Clinic of San Paolo Hospital of Milan and University of Siena). IOP was measured once with a Goldmann applanation tonometer (GAT) and twice by ART (ART1, ART2) in a randomized sequence, by a single operator to assess intra-operator variability. To assess inter-operator variability 2 operators, chosen randomly (A and B, B and C, A and C), performed 2 measurements each (in random order; i.e. A B A B; B A B A) on the same patient. Repeatability and reproducibility were assessed by the coefficient of variation (CoV) and intraclass correlation coefficient (ICC). Results:ART1 was 0.4±2.8 mmHg higher than ART2 (p<0.001). Results were not influenced by randomization test order. Intra-operator CoV was 7.6%±7.6%, and ICC was 0.88 (95%CI: 0.85-0.91). Inter-operator CoV ranged between 6.0%±10.8% and 8.0%±6.8%, and ICC between 0.906 and 0.917. ART1 and 2 were respectively 1.8±3.8 and 1.4±4.3 mmHg higher than GAT (p<0.001). Results were unaffected by diagnosis, age, central corneal thickness, keratometry, operator, randomization sequence. Conclusions:ART repeatability and reproducibility were almost perfect. Tonometry measurements with ART tended to be overestimated compared with GAT.

Repeatibility and reproducibility of applanation resonance tonometry / P. Fogagnolo, P. Frezzotti, P. Mittica, M. Digiuni, R. Paderni, E. Vallenzasca, L. Rossetti. - In: INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE. - ISSN 0146-0404. - 53:(2012 Mar 26), pp. E5042-E5042. ((Intervento presentato al convegno Association for Research in Vision and Ophthalmology tenutosi a Fort Lauerdale nel 2012.

Repeatibility and reproducibility of applanation resonance tonometry

P. Fogagnolo
Primo
;
E. Vallenzasca
Penultimo
;
L. Rossetti
Ultimo
2012

Abstract

Purpose:To assess repeatability (intra-observer variability) and reproducibility (inter-observer variability) of intraocular pressure (IOP) measurements with Applanation Resonance Tonometry (ART) and to evaluate possible influential factors. Methods:The study included 178 patients (101 glaucoma, 63 healthy and 14 with OHT; one eye per subject) from two Italian centers (Eye Clinic of San Paolo Hospital of Milan and University of Siena). IOP was measured once with a Goldmann applanation tonometer (GAT) and twice by ART (ART1, ART2) in a randomized sequence, by a single operator to assess intra-operator variability. To assess inter-operator variability 2 operators, chosen randomly (A and B, B and C, A and C), performed 2 measurements each (in random order; i.e. A B A B; B A B A) on the same patient. Repeatability and reproducibility were assessed by the coefficient of variation (CoV) and intraclass correlation coefficient (ICC). Results:ART1 was 0.4±2.8 mmHg higher than ART2 (p<0.001). Results were not influenced by randomization test order. Intra-operator CoV was 7.6%±7.6%, and ICC was 0.88 (95%CI: 0.85-0.91). Inter-operator CoV ranged between 6.0%±10.8% and 8.0%±6.8%, and ICC between 0.906 and 0.917. ART1 and 2 were respectively 1.8±3.8 and 1.4±4.3 mmHg higher than GAT (p<0.001). Results were unaffected by diagnosis, age, central corneal thickness, keratometry, operator, randomization sequence. Conclusions:ART repeatability and reproducibility were almost perfect. Tonometry measurements with ART tended to be overestimated compared with GAT.
intraocular pressure
Settore MED/30 - Malattie Apparato Visivo
26-mar-2012
http://abstracts.iovs.org//cgi/content/abstract/53/6/5042?sid=d42fcbd0-b6e7-4966-9bf0-08ed425ffbd0
Article (author)
File in questo prodotto:
Non ci sono file associati a questo prodotto.
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/207347
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus ND
  • ???jsp.display-item.citation.isi??? ND
social impact