A procedure for using citrated fresh plasmas for International Sensitivity Index (ISI) calibration of two types of whole blood point-of-care test (POCT) prothrombin time (PT) monitor systems has been assessed in a multicentre study. The CoaguChek Mini and TAS PT-NC systems gave higher ISI with whole blood samples than with fresh plasma calibrations. However, there was good agreement between whole blood and fresh plasma monitor system International Normalised Ratio (INR) and the reference INR of target samples. Reliable INR can therefore be obtained with both whole blood and plasma samples on these two POCT systems based on their respective ISI. With the CoaguChek Mini system, the plasma calibration ISI can also be used to derive reliable INR with whole blood PT results. This was not possible with the TAS PT-NC system.

European Concerted Action on Anticoagulation (ECAA). Comparison of fresh plasma and whole blood multicentre ISI calibrations of Coaguchek Mini and TAS PT-NC whole blood prothrombin time point-of-care monitors / L. Poller, M. Keown, N. Chauhan, A.M.H.P. van den Besselaar, A. Tripodi, C. Shiach, J Jespersen. - In: THROMBOSIS AND HAEMOSTASIS. - ISSN 0340-6245. - 87:5(2002), pp. 859-866.

European Concerted Action on Anticoagulation (ECAA). Comparison of fresh plasma and whole blood multicentre ISI calibrations of Coaguchek Mini and TAS PT-NC whole blood prothrombin time point-of-care monitors

A. Tripodi;
2002

Abstract

A procedure for using citrated fresh plasmas for International Sensitivity Index (ISI) calibration of two types of whole blood point-of-care test (POCT) prothrombin time (PT) monitor systems has been assessed in a multicentre study. The CoaguChek Mini and TAS PT-NC systems gave higher ISI with whole blood samples than with fresh plasma calibrations. However, there was good agreement between whole blood and fresh plasma monitor system International Normalised Ratio (INR) and the reference INR of target samples. Reliable INR can therefore be obtained with both whole blood and plasma samples on these two POCT systems based on their respective ISI. With the CoaguChek Mini system, the plasma calibration ISI can also be used to derive reliable INR with whole blood PT results. This was not possible with the TAS PT-NC system.
Settore BIO/12 - Biochimica Clinica e Biologia Molecolare Clinica
Settore MED/09 - Medicina Interna
Settore MED/15 - Malattie del Sangue
2002
Article (author)
File in questo prodotto:
Non ci sono file associati a questo prodotto.
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/189910
Citazioni
  • ???jsp.display-item.citation.pmc??? 2
  • Scopus 17
  • ???jsp.display-item.citation.isi??? ND
social impact