International sensitivity index calibrations based on the W.H.O. recommended method depend on orthogonal regression analysis. As this is not readily available in statistical packages, comparison has been made with simple linear regression analysis in a study of coagulometer effects on the International Normalized Ratio (INR) at 155 European centres. Sets of seven lyophilized normal and 20 lyophilized artificially depleted abnormal plasmas were provided with five coumarin test plasmas and two European Concerted Action on Anticoagulation reference thromboplastins (low International Sensitivity Index (ISI) human and high ISI rabbit). Local ISI based on the artificially depleted lyophilized plasmas using conventional orthogonal regression gave good correction for local coagulometer effects on the human reagent and minimal correction with the rabbit reagent INR. Results were considerably worse after attempts at correction using calibration based on linear regression analysis with both reagents. The results indicate that calibration of coagulometer prothrombin time systems using simple linear regression is not appropriate.

A comparison of linear and orthogonal regression analysis for local INR determination in ECAA coagulometer studies / L. Poller, A.M.H.P. van den Besselaar, J. Jespersen, A. Tripodi, D. Houghton. - In: BRITISH JOURNAL OF HAEMATOLOGY. - ISSN 0007-1048. - 102:4(1998), pp. 910-917. [10.1046/j.1365-2141.1998.00866.x]

A comparison of linear and orthogonal regression analysis for local INR determination in ECAA coagulometer studies

A. Tripodi
Penultimo
;
1998

Abstract

International sensitivity index calibrations based on the W.H.O. recommended method depend on orthogonal regression analysis. As this is not readily available in statistical packages, comparison has been made with simple linear regression analysis in a study of coagulometer effects on the International Normalized Ratio (INR) at 155 European centres. Sets of seven lyophilized normal and 20 lyophilized artificially depleted abnormal plasmas were provided with five coumarin test plasmas and two European Concerted Action on Anticoagulation reference thromboplastins (low International Sensitivity Index (ISI) human and high ISI rabbit). Local ISI based on the artificially depleted lyophilized plasmas using conventional orthogonal regression gave good correction for local coagulometer effects on the human reagent and minimal correction with the rabbit reagent INR. Results were considerably worse after attempts at correction using calibration based on linear regression analysis with both reagents. The results indicate that calibration of coagulometer prothrombin time systems using simple linear regression is not appropriate.
Coagulometers; INR; Linear; Orthogonal; Regression
Settore AGR/16 - Microbiologia Agraria
Settore MED/09 - Medicina Interna
Settore MED/15 - Malattie del Sangue
1998
Article (author)
File in questo prodotto:
Non ci sono file associati a questo prodotto.
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/189622
Citazioni
  • ???jsp.display-item.citation.pmc??? 2
  • Scopus 11
  • ???jsp.display-item.citation.isi??? 13
social impact