Whole blood coagulation monitors are increasingly used for patient self-testing to control oral anticoagulation, but there are no comprehensive quality assurance (QA) programs to check their performance. We report on the experience with one of such programs applied in a field study where patients on prothrombin time (PT)-international normalized ratio (INR) self-testing were asked to bring their monitors to the anticoagulation clinic for checking. PT-INR testing was performed three times over 3 months with 14 patient's monitors and test strips on three recalcified QA plasmas by an experienced laboratory operator. Each patient was also asked to perform PT-INR self-testing (his/her own capillary blood) which was then compared to the laboratory PT-INR (plasma). Overall, the comparison between the observed and the consensus PT-INR on QA plasmas was acceptable with the majority of measurements lying within ±15% or 20% of the consensus values. The comparison between the PT-INR self-testing and the laboratory method was also acceptable: overall, there was no statistical significant difference between the mean PT-INR values and the majority of paired measurements were less than 15% or 20% apart. In conclusion, our results show that the proposed QA scheme is feasible and may be implemented on a larger scale. Monitors should be recalled periodically to the clinic where they have been prescribed to the patient. During each visit, the clinic may check the monitors and patient self-testing performance as described. Such comprehensive QA system would make monitoring of oral anticoagulant treatment by self-testing safer and more effective.

Quality Assurance (QA) Program for Whole Blood Prothrombin Time (PT) International Normalized Ratio (INR) Point-of-Care Monitors Used for Patient Self-testing to Control Oral Anticoagulation / A. Tripodi, C. Bressi, M. Carpenedo, V. Chantarangkul, M. Clerici, P.M. Mannucci. - In: THROMBOSIS RESEARCH. - ISSN 0049-3848. - 113:1(2004), pp. 35-40.

Quality Assurance (QA) Program for Whole Blood Prothrombin Time (PT) International Normalized Ratio (INR) Point-of-Care Monitors Used for Patient Self-testing to Control Oral Anticoagulation

A. Tripodi;
2004

Abstract

Whole blood coagulation monitors are increasingly used for patient self-testing to control oral anticoagulation, but there are no comprehensive quality assurance (QA) programs to check their performance. We report on the experience with one of such programs applied in a field study where patients on prothrombin time (PT)-international normalized ratio (INR) self-testing were asked to bring their monitors to the anticoagulation clinic for checking. PT-INR testing was performed three times over 3 months with 14 patient's monitors and test strips on three recalcified QA plasmas by an experienced laboratory operator. Each patient was also asked to perform PT-INR self-testing (his/her own capillary blood) which was then compared to the laboratory PT-INR (plasma). Overall, the comparison between the observed and the consensus PT-INR on QA plasmas was acceptable with the majority of measurements lying within ±15% or 20% of the consensus values. The comparison between the PT-INR self-testing and the laboratory method was also acceptable: overall, there was no statistical significant difference between the mean PT-INR values and the majority of paired measurements were less than 15% or 20% apart. In conclusion, our results show that the proposed QA scheme is feasible and may be implemented on a larger scale. Monitors should be recalled periodically to the clinic where they have been prescribed to the patient. During each visit, the clinic may check the monitors and patient self-testing performance as described. Such comprehensive QA system would make monitoring of oral anticoagulant treatment by self-testing safer and more effective.
Settore BIO/12 - Biochimica Clinica e Biologia Molecolare Clinica
Settore MED/09 - Medicina Interna
Settore MED/15 - Malattie del Sangue
2004
Article (author)
File in questo prodotto:
Non ci sono file associati a questo prodotto.
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/189189
Citazioni
  • ???jsp.display-item.citation.pmc??? 0
  • Scopus 18
  • ???jsp.display-item.citation.isi??? 14
social impact