Background. To standardize the measurement of human serum proteins, a new reference material (RM)(IRMM ERM-DA 470k/ IFCC) has recently been released. We evaluatedthe commutability of this RM for two immunochemical albumin methods based on turbidimetry (Roche Cobas Tina-quant Gen. 2) and nephelometry (Delta Radim) principles. Methods. We measured albumin concentrations on the two systems in20 human serum samples (albumin concentrations, 16.3 to 42.3 g/L) and in the RM in duplicate in two different runs on the same day. Biological samples were collected, aliquoted and stored(£24 h) at +4°C until their use. Manufacturer’s control materials were used to validate analytical runs. The RM commutability was estimated from data of regression analysis using the 95% prediction interval (95PI) and multiples of the standard error of regression (Sy-x), according to the CLSI C53-A document. Results. The RM results were contained within the 95PIbasedon the results for the native clinical samples. In addition, using an acceptance criterion for commutability of ±2 times the experimental Sy-x (±3.135), the relative residual for the RM (-2.107) was within the acceptable range. Albumin values for RM on both analytical systems (34.6 g/L for turbidimetry and 34.2 g/L for nephelometry) were, however, markedly lower than expected (certified value±uncertainty, 37.2±1.2 g/L). Conclusions. The new RM is commutable between the evaluated immunochemical assays and can be used as a basis to maintain their traceability to higher-order references. However, some inconsistency in the assignment of values to the manufacturer’s working calibrators is likely and should be verified.
Evaluation of commutability of the ERM - DA 470k reference material for two albumin assays using immunochemical principles / I. Infusino, F. Braga, C. Valente, M. Panteghini. - In: CLINICAL CHEMISTRY AND LABORATORY MEDICINE. - ISSN 1434-6621. - 49:Suppl.1(2011 May), pp. S805-S805. ((Intervento presentato al convegno IFCC - WordLab - EuroMedLab tenutosi a Berlin nel 2011 [10.1515/CCLM.2011.530].
Evaluation of commutability of the ERM - DA 470k reference material for two albumin assays using immunochemical principles
F. Braga;M. PanteghiniUltimo
2011
Abstract
Background. To standardize the measurement of human serum proteins, a new reference material (RM)(IRMM ERM-DA 470k/ IFCC) has recently been released. We evaluatedthe commutability of this RM for two immunochemical albumin methods based on turbidimetry (Roche Cobas Tina-quant Gen. 2) and nephelometry (Delta Radim) principles. Methods. We measured albumin concentrations on the two systems in20 human serum samples (albumin concentrations, 16.3 to 42.3 g/L) and in the RM in duplicate in two different runs on the same day. Biological samples were collected, aliquoted and stored(£24 h) at +4°C until their use. Manufacturer’s control materials were used to validate analytical runs. The RM commutability was estimated from data of regression analysis using the 95% prediction interval (95PI) and multiples of the standard error of regression (Sy-x), according to the CLSI C53-A document. Results. The RM results were contained within the 95PIbasedon the results for the native clinical samples. In addition, using an acceptance criterion for commutability of ±2 times the experimental Sy-x (±3.135), the relative residual for the RM (-2.107) was within the acceptable range. Albumin values for RM on both analytical systems (34.6 g/L for turbidimetry and 34.2 g/L for nephelometry) were, however, markedly lower than expected (certified value±uncertainty, 37.2±1.2 g/L). Conclusions. The new RM is commutable between the evaluated immunochemical assays and can be used as a basis to maintain their traceability to higher-order references. However, some inconsistency in the assignment of values to the manufacturer’s working calibrators is likely and should be verified.Pubblicazioni consigliate
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.