eThe measurement of both gamma dose and thermal neutron fluence in a BNCT gamma-neutron mixed field can be achieved by means of a single thermoluminescence dosimeter (TLD-700), exploiting the shape of the glow-curve (GC). The method is based on simple algorithms containing parameters obtained from the TLD-700 GC and requires the gamma calibration GC (for gamma dose measurement) or the thermal neutron calibration GC (for neutron fluence measurement) and moreover the GC of a TLD-600 exposed to a BNCT field, uncalibrated. Some results are reported, showing the potentiality of the method.
Determination of gamma dose and thermal neutron fluence in BNCT beams from the TLD-700 glow curve shape / G. Gambarini, G. Bartesaghi, S. Agosteo, E. Vanossi, M. Carrara, M. Borroni. - In: RADIATION MEASUREMENTS. - ISSN 1350-4487. - 45:3/6(2010), pp. 640-642. [10.1016/j.radmeas.2010.02.006]
Determination of gamma dose and thermal neutron fluence in BNCT beams from the TLD-700 glow curve shape
G. GambariniPrimo
;G. BartesaghiSecondo
;M. CarraraPenultimo
;
2010
Abstract
eThe measurement of both gamma dose and thermal neutron fluence in a BNCT gamma-neutron mixed field can be achieved by means of a single thermoluminescence dosimeter (TLD-700), exploiting the shape of the glow-curve (GC). The method is based on simple algorithms containing parameters obtained from the TLD-700 GC and requires the gamma calibration GC (for gamma dose measurement) or the thermal neutron calibration GC (for neutron fluence measurement) and moreover the GC of a TLD-600 exposed to a BNCT field, uncalibrated. Some results are reported, showing the potentiality of the method.Pubblicazioni consigliate
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.